• o Chromatography Systems <br /> The chromatographic systems are calibrated using either the <br /> external of the internal standard techniques . Specific methods <br /> may impose variations and/or different acceptance criteria upon <br /> these two techniques_ These specific requirements are found in <br /> the CHEMUEST SOP covering the particular method in question. <br /> 1 <br /> a External Standard Calibration Procedure= <br /> For each analyte, or group of analytes, flue concentration <br /> levels of standards are prepared by adding aliquots of one or <br /> more stock standards to ualumetric flasks and diluting to <br /> 1 volume with an appropriate solvent. One of the standards <br /> should be at a concentration near the method detection limit. <br /> The other concentrations should define the working range of <br /> the system . <br /> Each of the calibration standards is introduced into the <br /> chromatography system using the technique to be used for <br /> introduction of actual samples Cie_ 2- to S-ul liquid <br /> Injections, purge IL trap, etc _) . Both a Calibration Curue, <br /> and a series of Calibration Factors (Cr,$) at each standard <br /> concentration, is calculated for each analyte . <br /> o The Calibration Curve is a plat of amount introduced <br /> us . detector response . <br /> amount introduced <br /> o The CF - ------------------w------------ (for nultiresponse <br />' total response <br /> compounds use the total area of all peaks used for <br /> quantitation) . <br /> o If the percent relative standard deviation CXRSO) <br /> between the cr,s is less than 2OX over the working <br /> r range, linearity through the origin can be assumed and <br /> an average CF can be used for quantitation . <br /> o If the above criteria is not net, the Calibration <br /> Curue can be used for quantitation if the residual, <br /> r, is greater than O .99S . <br /> o If neither criteria is net, the standard analyses <br /> must be repeated if quantitation is to be performed . <br /> o If the quantitation criteria are not met, <br /> documentation of the ability to sec the required <br /> r minimum detectable concentration is sufficient to <br /> determine the presence or absence of target compounds. <br /> 18 <br />