DATA UHLIDATIOH DHO REPORT RPPROUHL PROCESS <br /> a Data Ualidation= <br /> Both the analysts and the managers of each laboratory section <br /> review the data generated in their laboratory to insure that it <br /> meets all of the applicable analytical and quality control <br /> requirements_ Before the data is released to the Project <br /> Manager and Document Control staff, the manager** review <br /> includes the following itenss <br /> o Check that the analysis was performed as specified by the <br /> client. <br /> N <br /> a Check that all of the required samples were analyzed. <br /> a Check that no contaminants were present in blanks at <br /> greater than the quality acceptance limits. <br /> a Check that the required number of blanks, spikes, <br /> duplicates, standards, etc . were analyzed. <br /> o Check that all surrogate and spike recoveries are within <br /> quality acceptance lxmita. <br /> o Uerify that instrument calibration is acceptable . <br /> a Uerify that instrument tuning is acceptable . <br /> o Uerify that internal standard responses are acceptable . <br /> a Check/uerify that any method/client specific analytical, or <br /> quality control parameters are acceptable _ <br /> a Document any problems or exceptions to the above criteria <br /> (le . quality Rasurance Notices) . <br /> 1 <br /> 1 <br /> 1� <br /> 29 <br />