Laserfiche WebLink
r1seco <br /> � A Corning Company <br /> I Sample Description <br /> See the attached Sample Description Information <br /> The samples were received under chain-of-custody. <br /> II Analysis Reguest <br /> The following analytical tests were requested. <br /> Lab ID Analysis Description <br /> 060854-1 through 7 Total Petroleum Hydrocarbon (Gasoline) and BTEX <br /> -8 through 10 Nitrate <br /> III Quality Control <br /> A. Pro.iect Specific QC. No project specific QC (i .e , spikes and/or <br /> duplicates) was requested. <br /> B. Method Blank Results. A method blank is a laboratory-generated <br /> sample which assesses the degree to which laboratory operations <br /> and procedures cause false-positive analytical results for your <br /> samples <br /> No target parameters were detected in the method blanks associated <br /> with your samples at the reporting limit levels noted on the <br /> attached Method Blank Report. <br /> C. Laboratory Control Samples - The LCS Program <br /> DuplicateControl Samples. A DCS is a well-characterized matrix <br /> (blank water, sand or celite) which is spiked with certain target <br /> parameters and analyzed at approximately 10% of the sample load in <br /> order to establish method-specific control limits The DCS <br /> results associated with your samples are on the attached Duplicate <br /> Control Sample Report. <br /> Accuracy is measured by Percent Recovery as in: <br /> % recovery = measured concentration x 100 <br /> (actual concentration) <br /> Precision is measured using duplicate tests by Relative Percent Difference <br /> (RPD) as in: <br /> RPD = % recovery test I - % recovery test 2) x 100 <br /> (% recovery test I + % recovery test 2)/2 <br />