Laserfiche WebLink
Summary of Findings: <br /> All samples were analyzed according to the accompanying chains of custody. All analytical holding times <br /> were met. <br /> QC data were reviewed for laboratory and instrument precision and accuracy from LCS/LCSD recoveries <br /> and relative percent differences (RPDs) and MS/MSD sample recoveries and RPDs. All quality control <br /> elements were found to be within acceptable limits with the following exceptions: <br /> • The MS/MSD percent recoveries and RPD associated with the sulfate analysis of MW-101 and <br /> MW-201 were not meaningful because the concentration of sulfate in the parent samples was <br /> greater than four times the concentration used for the spike. Therefore, the criteria were not <br /> evaluated. <br /> • The MS/MSD percent recoveries associated with the metals analysis of TW-13 were less than the <br /> lower control limit for iron. Iron was detected in the sample and was qualified as estimated (J). <br /> Additionally the MS/MSD percent recoveries and RPD were not meaningful for manganese <br /> because the concentration in the parent sample was greater than four times the concentration <br /> used for the spike. Therefore, the criteria were not evaluated. <br /> • The MS/MSD percent recoveries and RPDs associated with the metals analysis of TW-17 were <br /> not meaningful for iron and manganese because the concentration in the parent sample was <br /> greater than four times the concentration used for the spike. Therefore, the criteria were not <br /> evaluated. <br /> Field sampling precision was also evaluated by using the calculated RPD between results reported for the <br /> field duplicate pairs, which are listed above. All RPD results were found to be within the acceptable limits <br /> for precision for all methods with the following exception: <br /> • The field duplicate pair MW-448D and MW-448D Dup had an RPD that was greater than the <br /> control limit for managanese; the result was qualified as estimated (J) in the parent sample. <br /> No target analytes were detected in any method blank or field QC samples with the following exceptions: <br /> • Equipment Rinse and Field Blank, collected on July 19, 2011, had TOC detected at 0.3 mg/L <br /> and sulfate detected at 0.4 mg/L and 0.3 mg/L, respectively. Sulfate was detected in all <br /> associated samples at concentrations that were greater than five times the blank concentration <br /> and was not qualified. Seven samples had TOC detected below the reporting limit and less than <br /> five times the blank result; the sample results were qualified as not detected at the reporting limit <br /> (U). The remaining samples had TOC detected at concentrations that were greater than five <br /> times the blank concentration and were not qualified. <br /> • Equipment Rinse and Field Blank, collected on July 20, 2011, had TOC detected at 0.9 mg/L <br /> and TDS detected at 8.0 mg/L and 7.0 mg/L, respectively. Additionally, Field Blank had <br /> 1.2 mg/L of sulfate detected. TDS and sulfate were detected in all associated samples at <br /> concentrations that were greater than five times the blank concentrations and were not qualified. <br /> Eight samples had TOC detected below the reporting limit and less than five times the blank <br /> result; the sample results were qualified as not detected at the reporting limit (U). Twelve <br /> samples had TOC detected at concentrations that were greater than the reporting limit and less <br /> than five times the blank result; the results were qualified as estimated (J). The remaining <br /> samples had TOC detected at concentrations that were greater than five times the blank <br /> concentration and were not qualified. <br /> Page 2of6 <br />