Laserfiche WebLink
Summary of Findings: <br /> All samples were analyzed according to the accompanying chains of custody. All analytical holding times <br /> were met. <br /> QC data were reviewed for laboratory and instrument precision and accuracy from LCS/LCSD recoveries <br /> and relative percent differences (RPDs) and MS/MSD sample recoveries and RPDs. All quality control <br /> elements were found to be within acceptable limits with the following exceptions: <br /> • The MS/MSD percent recoveries and RPD associated with the sulfate analysis of 1-11 were not <br /> meaningful because the concentration of chromium in the parent sample was greater than four <br /> times the concentration used for the spike. Therefore, the criteria were not evaluated. <br /> • The MS/MSD percent recoveries associated with the metals analysis of MW-301 were less than <br /> the lower control limit for chromium and iron. Both metals were detected in the sample and were <br /> qualified as estimated (J). Additionally the MS percent recovery was greater than the upper <br /> control limit for sulfate. Sulfate was detected in the sample and was qualified as estimated (J). <br /> • The MS/MSD percent recoveries associated with the metals analysis of 1-11 and 1-11 Dup were <br /> greater than the upper control limit for sulfate. Sulfate was detected in the sample and was <br /> qualified as estimated (J). Additionally, the percent recoveries and RPD were not meaningful for <br /> chromium because the concentration in the parent sample was greater than four times the <br /> concentration used for the spike. Therefore, this criterion was not evaluated. <br /> Field sampling precision was also evaluated by using the calculated RPD between results reported for the <br /> field duplicate pairs, which are listed above. All RPD results were found to be within the acceptable limits <br /> for precision for all methods with the following exception: <br /> • The field duplicate pair 1-11 and 1-11 Dup had an RPD that was greater than the control limit for <br /> TOC; the result was qualified as estimated (J) in the parent sample. <br /> No target analytes were detected in any method blank or field QC samples with the following exceptions: <br /> • Equipment Rinse and Field Blank, collected on October 12, 2011, had TOC detected at 0.2 <br /> mg/L and 0.1 mg/L, respectively. Additionally, sulfate was detected in Equipment Rinse at 0.8 <br /> mg/L. Sulfate was detected in all associated samples at concentrations that were greater than <br /> five times the blank concentration and was not qualified. TW-13 had TOC detected below the <br /> reporting limit and less than five times the blank result; the sample result was qualified as not <br /> detected at the reporting limit (U). The remaining samples had TOC detected at concentrations <br /> that were greater than five times the blank concentration and were not qualified. <br /> • Equipment Rinse and Field Blank, collected on October 13, 2011, had TOC detected at 0.1 <br /> mg/L 0.07 mg/L, respectively. TW-16 had TOC detected below the reporting limit and less than <br /> five times the blank result; the result was qualified as not detected at the reporting limit (U). The <br /> remaining samples had TOC detected at concentrations that were greater than five times the <br /> blank concentration and were not qualified. <br /> • Equipment Rinse and Field Blank, collected on October 17, 2011, had TOC detected at 0.2 <br /> mg/L. MW-447D, MW-447S, MW-448D and MW-448S had TOC detected below the reporting <br /> limit and less than five times the blank result; the result was qualified as not detected at the <br /> reporting limit(U). <br /> All data submitted for this project are of known and acceptable quality as qualified, based on laboratory- <br /> established control limits and the data quality objectives. These data are considered acceptable for their <br /> intended purposes. <br /> Page 2 of 5 <br />