Laserfiche WebLink
Summary of Findings: <br /> All samples were analyzed according to the accompanying chains of custody and all analytical holding <br /> times were met. <br /> QC data were reviewed for laboratory and instrument precision and accuracy from LCS/LCSD recoveries <br /> and relative percent differences (RPDs) and MS/MSD sample recoveries and RPDs. All quality control <br /> elements were found to be within acceptable limits with the following exceptions: <br /> • The MS/MSD percent recoveries associated with the analysis of MW-302 were greater than the <br /> upper control limit for sulfate; the result was qualified as estimated (J) in the parent sample. <br /> • The MS/MSD percent recoveries and RPDs associated with the metals analysis of EW-1 and <br /> MW-205 (S-7) were not meaningful for manganese and chromium, respectively. The <br /> concentration of manganese and chromium in the parent samples was greater than four times the <br /> concentration used for the spike. Therefore, the criteria were not evaluated. <br /> • The MS/MSD percent recoveries and RPD associated with the analysis of MW-322 Dup were not <br /> meaningful for sulfate. The concentration of sulfate in the parent sample was greater than four <br /> times the concentration used for the spike. Therefore, the criteria were not evaluated. <br /> Field sampling precision was also evaluated by using the calculated RPD between results reported for the <br /> field duplicate pairs, which are listed above. All RPD results were found to be within the acceptable limits <br /> for precision for all methods with the following exception: <br /> • The field duplicate pair MW-322 and MW-322 Dup had RPDs that were greater than the control <br /> limit for arsenic and TOC; the results were qualified as estimated (J) in the parent sample. <br /> No target analytes were detected in any method blank or field QC samples with the following exceptions: <br /> • Equipment Rinse and Field Blank collected on January 24, 2012 were associated with the <br /> analysis of three samples and had TOC detected at 0.8 and 0.7 mg/L, respectively. TW-17 had <br /> TOC detected below the reporting limit and less than five times the blank result; the sample result <br /> was qualified as not detected at the reporting limit (U). TW-16 and TW-16 Dup had TOC <br /> detected at concentrations that were greater than the reporting limit, but less than five times the <br /> blank concentration, and were qualified as estimated (J). <br /> • One method blank associated with the analysis of nineteen samples collected January 24, 2012 <br /> had 7.0 mg/L of TDS detected. QC sample Equipment Rinse had TDS detected below the <br /> reporting limit and less than five times the blank result; the samples results was qualified as not <br /> detected at the reporting limit (U). The remaining samples had TDS detected at concentrations <br /> that were greater than five times the method blank result, so no data were qualified. <br /> • Equipment Rinse and Field Blank collected on January 25, 2012 were associated with the <br /> analysis of two samples and had TOC detected at 0.2 and 0.3 mg/L, respectively. Additionally, <br /> Equipment Rinse had 5.0 mg/L of TDS detected and 0.47 mg/L of sulfate detected. All three <br /> constituents were detected in the associated samples at concentrations that were greater than <br /> five times the blank concentration and were not qualified. <br /> • Equipment Rinse and Field Blank collected on January 26, 2012 were associated with the <br /> analysis of five samples had 0.3 mg/L TOC detected. MW-447D, MW-447S, MW-448D, and MW- <br /> 447D Dup had TOC detected below the reporting limit and less than five times the blank result; <br /> the sample results were qualified as not detected at the reporting limit (U). MW-448S had TOC <br /> detected at a concentration that was greater than the reporting limit, but less than five times the <br /> blank concentration, and was qualified as estimated (J). <br /> Page 2 of 5 <br />