Laserfiche WebLink
Summary of Findings: <br /> All samples were analyzed according to the accompanying chains of custody and all analytical holding <br /> times were met. <br /> QC data were reviewed for laboratory and instrument precision and accuracy from LCS/LCSD recoveries <br /> and relative percent differences (RPDs) and MS/MSD sample recoveries and RPDs. All quality control <br /> elements were found to be within acceptable limits with the following exceptions: <br /> • The MS/MSD percent recoveries associated with the analysis of TW-15, TW-15 Dup, MW-321 (I- <br /> 9), and MW-321 (1-9) Duplicate were greater than the upper control limit for sulfate. Sulfate was <br /> detected in the samples and was qualified as estimated (J). <br /> • The MS/MSD percent recoveries and RPD associated with the analysis of MW-448D were not <br /> meaningful for sulfate. The concentration of sulfate in the parent samples was greater than four <br /> times the concentration used for the spike. Therefore, the criteria were not evaluated. <br /> Field sampling precision was also evaluated by using the calculated RPD between results reported for the <br /> field duplicate pairs, which are listed above. All RPD results were found to be within the acceptable limits <br /> for precision for all methods with the following exception: <br /> • The field duplicate pair MW-321 (1-9) and MW-321 (1-9) Duplicate had RPDs that were greater <br /> than the control limit for iron and manganese; the results were qualified as estimated (J) in the <br /> both samples. <br /> No target analytes were detected in any method blank or field QC samples with the following exceptions: <br /> • Equipment Rinse and Field Blank collected on April 10, 2013 were associated with the analysis <br /> of ten samples and had TOC detected 0.3 milligrams per liter (mg/L). Four samples had TOC <br /> detected below the reporting limit and less than five times the blank result, and were qualified as <br /> not detected (U) at the reporting limit. Three samples had TOC detected above the reporting <br /> limit, but less than five times the blank concentration, and were qualified as estimated (J). <br /> • Equipment Rinse and Field Blank collected on April 11, 2013 were associated with the analysis <br /> of seven samples TOC detected at 0.6 and 0.3 mg/L, respectively. Two associated sample had <br /> TOC detected below the reporting limit and less than five times the blank result; the sample <br /> results were qualified as not detected (U) at the reporting limit. Three samples had TOC detected <br /> at a concentration that was greater than the reporting limit, but less than five times the blank <br /> concentration, and were qualified as estimated (J). <br /> • Equipment Rinse and Field Blank collected on April 12, 2013 were associated with the analysis <br /> of sevem samples had TOC detected at 0.3 and 0.4 mg/L, respectively. Two samples had TOC <br /> detected below the reporting limit and less than five times the blank result; the sample results <br /> were qualified as not detected (U) at the reporting limit. Two samples had TOC detected at a <br /> concentration that was greater than the reporting limit, but less than five times the blank <br /> concentration, and were qualified as estimated (J). <br /> All data submitted for this project are of known and acceptable quality as qualified, based on laboratory- <br /> established control limits and the data quality objectives. These data are considered acceptable for their <br /> intended purposes. <br /> Page 2 of 4 <br />