Laserfiche WebLink
3LdLe VVaLCf Nesources t_ontrot tsoara rage z or L <br /> Liquid Sensor for Sumps 794390-206 *** <br /> Response time (min) 8.19 8.49 N/D**** <br /> Recovery time (min) <1 <1 N/D <br /> Product activation height(cm) 4.12 3.95 N/D**** <br /> Lower detection limit(cm) 4.67 4.36 N/D**** <br /> * ILS-250, TLS-250i, 250i Plus <br /> ** ILS-350, TLS-350 Series <br /> ***Liquid Sensor for Sumps 794390-206 is no longer manufactured and is not supported by the TLS-450 series <br /> console. It was replaced by sensor models 794380-208, 209 which are included on the NWGLDE list <br /> **** See Glossary <br /> *****This sensor is identified as an interstitial sensor, but will function properly in other containment areas such as <br /> containment sumps,contained risers and other contained enclosures. This sensor must be installed at the lowest point <br /> of the containment enclosure to insure early detection of liquids. <br /> Specificity Results(in addition to above): <br /> Activated: diesel, heating oil #2, water. <br /> Evaluations determined these sensors' responses to the liquids shown above and those indicated in the following <br /> section. Biodiesel blends 136-1320 meeting ASTM D7647 and biodiesel 8100 meeting ASTM D6751 would also <br /> produce an alarm if the sensor threshold is exceeded. Responses to these fuels were not determined, but would be <br /> expected to be very similar to the diesel responses. <br /> Comments: <br /> Comparison of TLS-450 Series console with TLS-350 series console showed comparable results when evaluated with <br /> representative sensors in water and diesel fuel. <br /> Sensors are reusable. <br /> Veeder-Root Evaluator: Carnegie Mellon Research Institute <br /> 125 Powder Forest Dr. Tel: (412) 268-3495 <br /> Simsbury, CT 06070-2003 Dates of Evaluation: 07/17/92, 04/22/98 <br /> Tel: (860)651-2700 <br /> E-mail: dhalla@veeder.com Evaluator: Ken Wilcox Associates <br /> URL: www.veeder.com/dynamic/index.cfm Tel: (816)443-2494 <br /> Dates of Evaluations: 3/3/2008 <br /> 1 Home I Vendors: A-F I Vendors: G - M I Vendors: N-S I Vendors: T-Z I Method Index I <br /> Appearance on this list is not to be construed as an endorsement by any regulatory agency nor is it any <br /> guarantee of the performance of the method or equipment. <br /> Equipment should be installed and operated in accordance with all applicable laws and regulations. <br /> For full details, please refer to our expanded "DISCLAIMER" page. <br /> Conditions of Use I Privacy Policy <br /> Copyright©2011 State of California <br /> The State Water Board is one of five boards,departments,and offices under <br /> the umbrella of the California Environmental Protection Agency. <br /> Cal/EPA I ARB I DPR I DTSC I OEHHA I SWRCB <br /> http://www.waterboards.ca.gov/water_issues/programs/ust/leak_prevention/lg 113/vendors/vee... 6/3/11 <br />