Laserfiche WebLink
l <br /> l <br /> iMATRIX SPIKE SAMPLE DATA REPORT <br /> GC/MS Volatiles <br /> Client Lot i# E1B130212 Work Order #. DV2001AL-ME Matrix_ _ . . . . . rPdER <br /> MS Lot-Sample #: EIB130212-001 DV2001AF-MSD <br /> Date Sampled_ : 02/07/01 09 08 Date Received. : 02/13/01 10 00 MS Run #- - - - - - - : 1045217 <br /> Prep Date_ . 02/14/01 Analysis Date_ : 02/14/01 <br />�I Prep Batch #. . . 1045462 Analysis Time. . : 18 38 <br /> Dilution Factor 25 <br /> SAMPLE SPIKE MEASRD PERCENT <br /> PARAMETER AMOUNT AMT AMOUNT UNITS RECOVERY RPD METHOD <br /> Benzene ND 250 258 ug/L 103 SW846 8260B <br /> ND 250 239 ug/L 95 7.9 SW846 8260B <br /> Chlorobenzene ND 250 266 ug/L 106 SW846 8260B <br /> ND 250 248 ug/L 99 7.0 SW846 8260B <br /> 1,1-Dachloroethene ND 250 251 ug/L 100 SW846 8260B <br /> ND 250 241 ug/L 96 4.1 SW846 8260E <br /> Toluene ND 250 255 ug/L 102 SW846 8260B <br /> ND 250 242 ug/L 97 5.5 SW846 8260B <br /> Trichloroethene ND 250 280 ug/L 112 SW846 8260B <br /> ND 250 261 ug/L 105 6.6 SW846 8260B <br /> PERCENT RECOVERY <br /> ANkURROGATE RECOVERY LIMITS <br /> IUFSromofluorobenzene 102 (75 - 120) <br /> 92 (75 - 120) <br /> 1,2-Dichloroethane-d4 110 (65 130) <br /> 96 (65 - 130) <br /> Toluene-d8 105 (80 130) <br /> 94 (80 - 130) <br /> NOTE(S) <br /> Calculations are performed before rounding to avoid round-off errors to calculated results <br /> Bold print denotes control parameters <br />'1 <br />