Laserfiche WebLink
r <br /> August 25, 2006 <br /> Neil Anderson & Associates Lab ID <br /> STK637174 <br /> Quality Control - Organic Customer : 3-9574 <br /> r <br /> Constituent Method <br /> Date/ID Type Units <br /> DBCP Conc. QC Data DQO Note <br /> 504.1 08/22/2006:A203 Blank <br /> ug/L ND <br /> LCS ug/L 0.2941 <0.01 <br /> MS 96.8% 70-130 <br /> (SP 608239-01) MSD ug/L 0.2941 96.6% 70-130 <br /> MSRPD �g�L 0.2941 98.2% 70-130 <br /> 1.6% 530 <br /> 504.1 0822/2006:A 02-CCV <br /> 03-CCV ug/L 10.00 115% 70-130 <br /> ug/L 10.00 112% 70-130 <br /> 1,3-Dibr0m0pr0pane-Surrogate 504.1 08/222006:A203 Blank <br /> LCS ug/L 0.5882 115% 70-130 <br /> MS ug/L 0.5882 114% 70-130 <br /> ug/L 0.5882 112% 70-130 <br /> (SP 608239-01) MSD <br /> ug/L 0.5882 111% 70-130 <br /> 504.1 08/22/2006:A 02-CC V ug/L 9.975 124% 70-130 <br /> Definition, 03-CCV ug/L 9.975 122% 70-130 <br /> Blank Method Blank - Prepared to verify that the preparation process is not contributing contamina <br /> LCS Laboratory Control Standard/Sample - Prepared tion to the samples. <br /> preto verify that the preparation process is not affecting analyte <br /> recovery. <br /> MS/MSD Matrix Spikes - A random sample is spiked with a known amount of analyte. The recoveries are an indication o <br /> how that sample matrix affects analyte recovery. <br /> CCB Continuing Calibration Blank - Analyzed to verify the instrument baseline is within criteria. <br /> CCV Continuing Calibration Verification - Analyzed to verify the instrument calibration is within criteria. <br /> ND Non-detect - Result was below the DQO listed for the analyte. <br /> DQO Data Quality Objective - This is the criteria against which the quality control data is compared. <br /> STK637174: Quality Control Page 2 <br />