Laserfiche WebLink
March 20,2019 Lab ID : STK1933512 <br /> Live Oak Geo Environmental Customer : 3-16608 <br /> Quality Control - Inorganic <br /> Constituent Method Date/ID Type Units Cone. QC Data DQO Note <br /> Wet Chem <br /> Nitrate+Nitrite as N 450ONO3F MS mg/L 28.18 37.1% 5-285 <br /> (SP 1903305-001) MSD mg/L 28.18 38.1% 5-285 <br /> MSRPD m /L 28.18 2.4% <30A <br /> 450ONO3F 03/13/19:203612JDD ICB mg/L 0.110 0.2 <br /> ICV mg/L 11.27 96.9% 90-110 <br /> CCB mg/L 0.103 0.2 <br /> CCV m /L 11.27 973% 90-110 <br /> Definition <br /> ICV :Initial Calibration Verification-Analyzed to verify the instrument calibration is within criteria. <br /> ICB :Initial Calibration Blank-Analyzed to verify the instrument baseline is within criteria. <br /> CCV :Continuing Calibration Verification-Analyzed to verify the instrument calibration is within criteria. <br /> CCB :Continuing Calibration Blank-Analyzed to verify the instrument baseline is within criteria. <br /> MS :Matrix Spikes-A random sample is spiked with a known amount of analyte.The recoveries are an indication of how that sample <br /> matrix affects analyte recovery. <br /> MSD :Matrix Spike Duplicate of MS/MSD pair-A random sample duplicate is spiked with a known amount of analyte.The recoveries <br /> are an indication of how that sample matrix affects analyte recovery. <br /> MSRPD :MS/MSD Relative Percent Difference(RPD)-The MS relative percent difference is an indication of precision for the preparation <br /> and analysis. <br /> O :Data Quality Objective-This is the criteria against which the quality control data is compared. <br /> Page 9 of 9 <br />