Laserfiche WebLink
Client Name: Groundxater Technologies bate: 01/04/1995 <br /> Client ACCt: 417 ELAP Cert: 1386 <br /> ® NET Sob No: 94.06188 Page: 15 <br /> Ref: ARCO Facility 434, 501 w. Kettleman, Lodi, GTI-GC-A94-5 <br /> CONTINUING CALIBRATION VERIFICATION STANDARD REPORT <br /> CCV CCV <br /> CCV Standard Standard Run <br /> Standard Amount Amount Acceptance Date Batch <br /> Parameter t Recovery Found ectad Criteria Units Analyzed Number <br /> TPH (Gas/BTXE,Liquid) <br /> as Gasoline 104.0 1.04 1.00 85 - 115 mg/L 11/30/1994 2435 <br /> Benzene 99.4 4.97 5.00 85 - 115 ug/L 12/30/1994 2435 <br /> Toluene 88.8 4.44 5.00 85 - 115 ug/L 12/30/1994 2435 <br /> Ethylbenzene 92.4 4.62 5.00 85 - 115 ug/L 12/30/1994 2435 <br /> Xylenes (Total) 87.3 13.1 15.0 85 - 115 ug/L 12/30/1994 2435 <br /> Bromofluorobenzene (SURR) 87.0 87 100 85 - 115 t Rea. 12/30/1994 2435 <br /> TPH (Gas/BTXE,Liquid) <br /> as Gasoline 114.0 1.14 1.00 85 - 115 mg/L 12/30/1994 2442 <br /> Benzene 91.2 4.56 5.00 85 - 115 ug/L 12/30/1994 2442 <br /> Toluene 97.6 4.88 5.00 85 - 115 ug/L 12/30/1994 2442 <br /> Ethylbenzene 108.2 5.41 5.00 85 - 115 ug/L _ 12/30/1994 2442 _ <br /> Xylenes (Total) 90.7 13.6 15.0 65 - 115 ug/L - 12/30/1994 2443 <br /> Bromofluorobenzene ISURR) 108.0 108 100 85 - 115 t Rec. 12/30/1994 2442 <br /> NOTE: Results apply only to the samples analyzed. Reproduction of this report is permitted only in its entirety. <br />