Laserfiche WebLink
Case Narrative for job: 720-J7408-1 <br /> Client: SECOR International, Inc. <br /> Date: 02/02/2007 <br /> Semi Volatiles GC Analysis <br /> Method Blank- Report, 1 OX <br /> The method blank for preparation batch 17537 contained a target compound above the reporting limit(RL). The <br /> associated samples contained detects for this analyte at concentrations greater than 1 OX the value found in the <br /> method blank,therefore, re-extraction and/or re-analysis of samples was not performed. <br /> Affected Items <br /> 720-7408-A-7-A <br /> Batch: 720-17745 <br /> Method: 720-80156 DRO <br /> Semi Volatiles GC Analysis <br /> Method Blank- Report, ND <br /> The method blank for preparation batch 17537 contained a target compound above the reporting limit(RL). None <br /> of the samples associated with this method blank contained the target compound; therefore, re-extraction and/or <br /> re-analysis of samples were not performed. <br /> Affected Items <br /> 720-7408-A-2-A <br /> Batch: 720-17745 <br /> Method: 720-8015 B_DRO <br /> 720-7408-A-4-A <br /> Batch: 720-17745 <br /> Method: 720-801513_DRO <br /> 720-7408-A-5-A <br /> Batch: 720-17745 <br /> Method: 720-8015B_DRO <br /> 720-7408-A-6-A <br /> Batch: 720-17745 <br /> Method: 720-8015 B_D RO <br /> 720-7408-A-8-A <br /> Batch: 720-17745 <br /> Method: 720-8015B_DRO <br /> 720-7408-A-10-A <br /> Batch: 720-17745 <br /> Method: 720-8015B DRO <br /> Volatiles MS <br /> Surrogate- Diluted out <br /> Due to the level of dilution required for sample 7408-a-7 , surrogate recoveries are not reported. <br /> Affected Items <br /> 720-7408-A-7-B <br /> Batch: 720-17644 <br /> Method: 720-8260B <br />