urfi <br /> Air Toxics <br /> LABORATORY NARRATIVE <br /> Modified TO-15 Low Level <br /> Advanced GeoEnvironmental <br /> Workorder# 1606238 <br /> Three 6 Liter Summa Canister (100% Certified) samples were received on June 10, 2016. The <br /> laboratory performed analysis via modified EPA Method TO-15 using GUMS in the full scan mode. <br /> This workorder was independently validated prior to submittal using 'USEPA National Functional <br /> Guidelines' as generally applied to the analysis of volatile organic compounds in air. A rules-based, <br /> logic driven, independent validation engine was employed to assess completeness, evaluate pass/fail of <br /> relevant project quality control requirements and verification of all quantified amounts. <br /> Method modifications taken to run these samples are summarized in the table below. Specific project <br /> requirements may over-ride the ATL modifications. <br /> Requirement TO-15 ATL Modifications <br /> Initial Calibration </=30%RSD with 2 </=30%RSD with 4 compounds allowed out to<40% <br /> compounds allowed out RSD <br /> to<40%RSD <br /> Blank and standards Zero Air UHP Nitrogen provides a higher purity gas matrix than <br /> zero air <br /> Receiving Notes <br /> The Chain of Custody (COC) information for sample IA-VW-rear did not match the information on the <br /> canister with regard to canister identification. The client was notified of the discrepancy and the <br /> information on the canister was used to process and report the sample(s). <br /> Analytical Notes <br /> Dilutions were performed on samples IA-VW-rear, IA-VW-front and IA-VW-show room due to the <br /> presence of high level target species. <br /> The reported result for 4-Ethyltoluene in samples IA-VW-rear, IA-VW-front and IA-VW-show room <br /> may be biased high due to co-elution with a non target compound with similar characteristic ions. Both <br /> the primary and secondary ion for 4-Ethyltoluene exhibited potential interference. <br /> All Quality Control Limit exceedances and affected sample results are noted by flags. Each flag is <br /> defined at the bottom of this Case Narrative and on each Sample Result Summary page. Target <br /> compound non-detects in the samples that are associated with high bias in QC analyses have not been <br /> Ragged. <br /> The Relative Percent Difference (RPD) of the LCS/LCSD exceeded acceptance limits for <br /> Bromomethane. <br /> Definition of Data Oualifyin2 Flags <br /> Eight qualifiers may have been used on the data analysis sheets and indicates as follows: <br /> B - Compound present in laboratory blank greater than reporting limit (background subtractior. <br /> Page 3 of 20 <br />