APEX EIVViROTECH, INC.
<br /> STANDARD OPERATING PROCEDURES
<br /> Quarterly Monitoring Reports
<br /> SOP - 4 When practical,the groundwater sample should be collected
<br /> SAMPLE IDENTIFICATION AND CHAIN- when the water level in the well recovers to at least 60 percent
<br /> OF-CUSTODY PROCEDURES of its static level
<br /> The sampling equipment consists of either a"Teflon"bailer,
<br /> Sample identification and chain-0f-custody procedures ensure PVC bailer,or stainless steel bladder pump with a"Teflon'
<br /> a sample integrity as well as document sample possession from bladder if the sampling system is dedicated to the well,then
<br /> the bme of collection to ultimate disposal Each sample the bailer Is usually"Teflon,"but the bladder pump is PVC with
<br /> container submitted for analysis is labeled to identify the job a polypropylene bladder In general and depending on the
<br /> number,date,time of sample collection,a sample number intended laboratory analysis,40-millilder glass,volatile organic
<br /> unique to the sample,any in-field measurements made, analysis(VOA)vials,with'Teflon"septa,are used as sample
<br /> sampling methodology,name(s)of on-site personnel,and any containers
<br /> other pertinent field observations also recorded an the Field
<br /> excavation or boring log SOP - 12
<br /> MEASURING LIgU1D LEVELS USING
<br /> Chain-of-custody forms are used to record possession of the WATER LEVEL METER OR INTERFACE
<br /> sample from time of collection to arrival at the laboratory PROBE
<br /> During shipment,the person with custody of the samples wilt
<br /> Field equipment used for liquid-level gauging typically includes
<br /> relinquish them to the next person by signing the chain-of-
<br /> the measuring instrument(water level meter or interface probe)
<br /> custody form(s)and noting the date and time The sample
<br /> control officer at the laboratory will verify sample Integrity, and product bailer(s) The field kit also includes cleaning
<br /> correct preservation,confirm collection In the proper supplies(buckets,solution,spray battles,and deianized water)
<br /> container(s),and ensure adequate volume for analysis to be used in cleaning the equipment between wells
<br /> If these conditions are met,the samples will be assigned Prior to measurement,the instrument tip is lowered into the well
<br /> unique laboratory log numbers for identification throughout until d touches bottom Using the previously established top-of-
<br /> analysis and reporting The log numbers will be recorded on casing or top-of-box(e,wellhead vault)point,the probe cord
<br /> (or halyard)Is marked and a measuring tape(graduated In
<br /> the maintained
<br /> the la forms and a the legally-required log book hundredths of a foot)is used to determine the distance
<br /> maintained in the laboratory The sample description.date between the probe end and the marking on the cord This
<br /> received, lient's name and any other relevant information will measurement is then recorded on the liquid-level data sheet as
<br /> also be reccorded the"Measured Total Depth"of the well
<br /> SOP - S When necessary in using the interface probe to measure liquid
<br /> LABORATORY ANALYTICAL QUALITY levels,the probe is first electncalty grounded to either the metal
<br /> ASSURANCE AND CONTROL stove pipe or another metal object nearby When no ground is
<br /> available,reproducible measurements can be obtained by
<br /> clipping the ground lead to the handle of the Interface probe
<br /> In addition to routine instrument calibration,replicates,spikes, case
<br /> blanks,spiked blanks,and certified reference materials are
<br /> routinely analyzed at method-specific frequencies to monitor The probe tip is then lowered into the well and submerged In
<br /> precision and bias Additional components of the laboratory the groundwater An oscillating(beeping)tone indicates the
<br /> Quality Assurance/Quality Control program include probe Is In water The probe is slowly raised until either the
<br /> I Participation in state and federal laboratory oscillating tone ceases or becomes a steady tone In either
<br /> accreditation/certification programs, case,this is the depth-to-water(DTW)Indication and the IDTW
<br /> measurement is made accordingly The steady tone indicates
<br /> U S EPA Performance floating liquid hydrocarbons(FLH) In this case,the probe Is
<br /> 2 Participation in both
<br /> Evaluation studies US and WP studies)and inter- slowly raised until the steady tone ceases This Is the depth-to-
<br /> product(DTP)Indication and the DTP measurement Is made
<br /> laboratory performance evaluation programs, accordingly
<br /> 3 Standard operating procedures describing routine The process of lowering and raising the probe must be
<br /> and periodic Instrument maintenance, repeated several times to ensure accurate measurements The
<br /> "Out-of-Control"1Corrective Action documentation DTW and DTP measurements are recorded on the riqufd-level
<br /> q data sheet When FLH are Indicated by the probe's response,
<br /> procedures,and, a product bailer is lowered partially through the FLH-water
<br /> 5 Muhl-level review of raw data and client reports Interface to confirm the FLH on the water surface and as
<br /> further Indication of the FLH thickness,particularly In cases
<br /> where the FLH layer Is quite thin This measurement Is
<br /> SOP - 7 recorded on the data sheet as"FLH thickness"
<br /> GROUNDWATER PURGING AND
<br /> SAMPLING In order to avoid cross-contamination of wells during the(squid-
<br /> Prior to water sampling,each well is purged by evacuating a level measurement process,wells are measured in the order of
<br /> "clean"to"dirty'(where such information Is available) In
<br /> minimum of three wetted well-casing volumes of groundwater addition,all measurement equipment is cleaned with solution
<br /> When required,purging will continue until either the discharge and thoroughly rinsed with deianized water before use,between
<br /> water temperature,conductivity,or pH stabilize,a maximum of measurements in respective wells,and at the completion of the
<br /> ten wetted-casing volumes of groundwater have been days use
<br /> recovered,or the well is bailed dry