Laserfiche WebLink
memory, and then subtracts that background from all measured concentration <br /> values,until the zero is updated again by the user. <br /> Although zeroing can be performed as often as desired (e.g.,before every run),in <br /> practice it should not be necessary to do so more than once-a-month or even less <br /> frequently, except if average particulate concentrations should exceed about 0.5 <br /> mg/M-3. <br /> 6.5.1 Zeroing the model pDR-1000AN <br /> Zeroing of the model pDR-1000AN requires a particle-free environment such as a <br /> clean room,clean bench,duct or area directly downstream of a NEPA filter,or the <br /> pDR-1000AN Z-Pouch(standard accessory). In some cases,a very clean,well air <br /> conditioned office may offer a sufficiently low particle concentration environment <br /> (i.e.,<_5µg/m3) for zeroing,as determined by another monitor (e.g.,ME <br /> DataRAM). <br /> To zero the model pDR-1000AN by means of its Z-Pouch,proceed as follows: <br /> • Wipe the outside surfaces of the pDR-1000AN to remove as much dust from <br /> those surfaces as possible before placing the instrument inside the Z-Pouch. <br /> • In a reasonably clean environment,open the zipper of the Z-Pouch and place <br /> the pDR-1000AN inside it. Close the zipper shut. <br /> • Open the small nipple on the Z-Pouch,and insert the fitting of the hand <br /> pump/in-line filter unit into the nipple. <br /> • Start pumping the hand-pump until the Z-Pouch begins to bulge, and proceed <br /> with the steps in Section 8.1,pressing the keys of the instrument through the <br /> wall of the Z-Pouch.Then continue pumping. <br /> • After completing the zeroing (step 2. of Section 8.1)procedure,open the Z- <br /> Pouch zipper and remove the pDR-1000AN. Close the zipper and flatten the Z- <br /> Pouch while plugging its nipple,in order to prevent dust contamination of the <br /> interior of the Z-Pouch. <br /> • The pDR-1000AN is now zeroed and ready for a measurement run. <br /> 6.5.2 Zeroing the model pDR-1200 <br /> To provide the particle-free air required to zero the pDR-1200, either of two <br /> methods can be used: a) place the instrument on a clean-air bench or in a clean <br /> room,or b)connect to the cyclone inlet the green zeroing filter cartridge supplied <br /> with the pDR-1200. In either case,proceed as follows: <br /> • After implementing either of the two methods, above,run the attached pump <br /> for at least one minute (e.g.,at 4 liters/minute), and then proceed as described in <br /> Section 8.1 of this instruction manual,while continuing to run the pump (or <br /> leaving the unit in the clean air environment). <br /> 14 <br />