The primary dust concentration reference value is obtained from the weight <br /> increase of the two filters due to the dust collected over a measured period of time, <br /> at a constant and known flow rate. The two master real-tune monitors are then <br /> adjusted to agree with the reference mass concentration value (obtained from <br /> averaging the measurements of the two gravimetric filters) to within±1%. <br /> Three primary,NIST traceable,measurements are involved in the determination of <br /> the reference mass concentration: the weight increment from the dust collected on <br /> the filter, the sampling flowrate, and the sampling time. Additional conditions that <br /> must be met are: a) suspended dust concentration uniformity at all sampling inlets <br /> of the mixing chamber;b) identical sample transport configurations leading to <br /> reference and instrument under calibration; and c) essentially 100% collection <br /> efficiency of filters used for gravimetric reference for the particle size range of the <br /> test dust. <br /> The test dust used for the MIE factory calibration of the personalDataRAM is SAE <br /> Fine (ISO Fine) supplied by Powder Technology,Inc. It has the following physical <br /> characteristics (as dispersed into the mixing chamber): <br /> • Mass median aerodynamic particle diameter: 2 to 3 µm <br /> • Geometric standard deviation of lognormal size distribution: 2.5 <br /> • Bulk density: 2.60 to 2.65 g/cm3 <br /> • Refractive index: 1.54 <br /> 13.2 Field Gravimetric Calibration <br /> If desired,the personalDataRAM can be calibrated gravimetrically for a particular <br /> aerosol(dust,smoke,mist, etc.) under field conditions (actual conditions of use). <br /> To effect such calibration in the particle environment of interest, proceed as <br /> indicated below. <br /> For field calibration of the model pDR-1000AN, a personal type filter sampler is <br /> placed side-by-side (collocated) to the pDR-1000AN to be calibrated, and the two <br /> units should be started simultaneously. For the model pDR-1200,its own filter and <br /> attached pump can be conveniently used for the same purpose. <br /> • Weigh and load into filter holder a fresh membrane filter. <br /> • Start pump. <br /> • Immediately turn on personalDataRAM and start a run such that the pump and <br /> the personalDataRAM are started nearly simultaneously. <br /> The duration of this comparison run should be sufficient to collect a mass of at least <br /> 1 mg on the reference filter (in order to permit accurate weighing of the collected <br /> mass by means of an analytical balance). The time-weighted average (TWA) <br /> reading of the personalDataRAM can be used to estimate the required sampling time <br /> to collect the above-mentioned mass on the filter.To estimate the required sampling <br /> time (ET as measured on the personalDataRAM) in minutes,read the TWA value <br /> 30 <br />