Laserfiche WebLink
Evaluation Results <br /> Note: If the test data can be presented in a more appropriate manner, the evaluator <br /> may select to present the information below in a data table, which can be attached to <br /> these forms. <br /> Table 1. Results of the Evaluation <br /> Product <br /> Parameter Brine <br /> Threshold - Lower Detection Limit (inches) 0.793 <br /> Precision - Standard Deviation (inches) 0.00184 <br /> Detection Time (seconds) <15 <br /> Fall Time (seconds) <15 <br /> Specificity-This sensor will respond to any liquid after its threshold is exceeded but <br /> testing was conducted only with brine <br /> Additional Limitations or Considerations- None <br /> > Safety Disclaimer: This test procedure only addresses the issue of the methods <br /> ability to respond to liquids. It does not test the equipment for safety hazards. <br /> Certification of Results <br /> I certify that the liquid level sensor was tested under conditions according to the <br /> vendor's operating instructions. I also certify that the evaluation was performed using <br /> methods described in the attached Alternative EPA Test Procedures for Liquid level <br /> sensors, and that the results presented above are those obtained during the evaluation. <br /> H. Kendall Wilcox Ken Wilcox Associates, Inc. <br /> (printed name) (organization performing evaluation) <br /> Grain Valley, MO 64029 <br /> (signature) (city,state,zip) <br /> May 18. 2004 ) 443-2494 <br /> Liquid Level Sensor-Results Form Page 2 of 2 <br />