Laserfiche WebLink
GROUNDWATER TECHNOLOGY, INC. <br />• STANDARD OPERATING PROCEDURE <br /> CONCERNING CHAIN OF CUSTODY <br /> SOP 11 <br /> 1 Samples must be maintained under custody until shipped or delivered to the laboratory The <br /> laboratory will then maintain custody A sample Is under custody If <br /> a) It Is in your possession <br /> b) It Is in your view after being In your possession <br /> c) You locked it up after It was in your possession <br /> d) It Is In a designated secure area <br /> 2 Custody of samples may be transferred from one person to another Each transferrer and recipient <br /> must date, sign and note the time on the chain-of-custody form <br /> 3 In shipping,the container must be sealed with tape,and bear the sender's signature across the area <br /> of bonding at the ends of the tape to prevent undetected tampering Each sampling jar should be <br /> taped and signed as well Scotch tape works well <br /> 4 Write "sealed by" and sign In the "Remarks" box at the bottom of the form before sealing the box <br /> Place form in a plastic bag and seal it Inside the box <br /> 5 The "REMARKS" section of the form is for documenting details such as <br /> a) Correlation of sample numbers If samples are split between labs <br /> b) QC numbers when lab Is logging In the samples <br /> C) Sample temperature and condition when received by lab <br /> d) Preservation notation <br /> e) pH of samples when opened for analysis (if acidified) <br /> f) Sampling observation or sampling problem <br /> 6 The chain-of-custody form should be Included Inside the shipping container A copy should be sent <br /> to the project manager <br /> 7 When the samples are received by the lab,the chain-of-custody form will be dated, signed, and the <br /> time noted by a laboratory representative The form will be retained In the laboratory files along with <br /> shipping bills and receipts <br /> 8 At the time of receipt of samples by the laboratory,the shipping container will be inspected and the <br /> sealing signature will be checked The samples will be Inspected for condition and bubbles, and <br /> the temperature of a representative sample container will be measured externally by a thermocouple <br /> probe(held tightly between two samples) and recorded The laboratory QC numbers will be placed <br /> on the labels, In the accession log,and on the chain-of-custody form If samples are acidified, their <br /> pH will be measured by narrow range pH paper at the time of opening for analysis All comments <br /> concerning procedures requiring handling of the samples will be dated and Initialed on the form by <br /> the laboratory person performing the procedure A copy of the completed chain-of-custody form <br /> with the comments on sample Integrity will be returned to the sampler <br /> GROUNDWATER <br /> �� TECHNOLOGY, INC <br />