Laserfiche WebLink
1 <br />17 <br />u <br />C <br />1 <br />�L] <br />i <br />u <br />11 <br />0 <br />Evaluation Results <br />Note: If the test data can be presented in a more appropriate manner, the evaluator may <br />select to present the information below in a data table, which can be attached to these <br />forms. <br />Specificity -This sensor will respond to any liquid after its threshold is exceeded but <br />testing was conducted only with brine <br />Additional Limitations or Considerations -.None <br />Is-ruelag 1017 • - •' <br />Certification of Results <br />I certify that the liquid level sensor was tested under conditions according to the <br />vendor's operating instructions. I also certify that the evaluation was performed using <br />methods described in the attached Alternative EPA Test Procedures for Liquid level <br />sensors, and that the results presented above are those obtained during the evaluation. <br />H. Kendall Wilcox <br />(printed name) <br />(signature) <br />M <br />Liquid Level Sensor - Results Form <br />Ken Wilcox Associates, Inc. <br />(organization performing evaluation) <br />Grain Valley, MO 64029 <br />(city, state, zip) <br />(816) 443-2494 <br />Page 2 of 2 <br />Product <br />Parameter <br />Brine <br />Threshold - Lower Detection Limit (inches) <br />0.793 <br />Precision - Standard Deviation (inches) <br />0.00184 <br />Detection Time (seconds) <br /><15 <br />Fall Time (seconds) <br /><15 <br />Specificity -This sensor will respond to any liquid after its threshold is exceeded but <br />testing was conducted only with brine <br />Additional Limitations or Considerations -.None <br />Is-ruelag 1017 • - •' <br />Certification of Results <br />I certify that the liquid level sensor was tested under conditions according to the <br />vendor's operating instructions. I also certify that the evaluation was performed using <br />methods described in the attached Alternative EPA Test Procedures for Liquid level <br />sensors, and that the results presented above are those obtained during the evaluation. <br />H. Kendall Wilcox <br />(printed name) <br />(signature) <br />M <br />Liquid Level Sensor - Results Form <br />Ken Wilcox Associates, Inc. <br />(organization performing evaluation) <br />Grain Valley, MO 64029 <br />(city, state, zip) <br />(816) 443-2494 <br />Page 2 of 2 <br />