Laserfiche WebLink
X r**;UU #J— Guixi City Jinlong Tattoo Equipment Manufactory <br /> -Vt- -q File Number: JLTS-1908 T<jW6A—, Chamber No.: <br /> �iZI 94 p M Test Date: 19-08-18 <br /> Original Record of BL Growth Test after Sterilization <br /> Customer: Bishop Tattoo Supply <br /> I�ruu�sh Product Name: DA VINCI Cartridges <br /> I�nu#t "5 Product Lot No.:08-19 <br /> XX Q M Sterilization Date: 19-08-17Sterilization Lot.: 2019081701 <br /> ttfflklt Test Quantity: 20 pcs # 9 R Test Item: BI Growth Test Record <br /> Q M Test Date: 19-08-18 $M46-f4i= Test Standard: ISO 11737-2:2009 <br /> ttj9;-:#j-rkM fe- iel:R* BI Growth Test Record Time Table <br /> Strain: Bacillus Atrophaeus ATCC9372 BI Manufacturer: Hangzhou Future Biotech Co., Ltd <br /> BI Lot No.: 190818 i9MiE Incubation Tempearture: 36 °C <br /> M r,to-q *;jWfH]R-K!WR Culture Time and BI Growth <br /> Strip No.: 8 16 24 32 40 48 <br /> fp,�V A 1-10# -- -- -- — -- _- <br /> [—Tl# + + + + + + <br /> A Result: (1-10#) �E*rFJR)EMI-K, MitkIM (11#) 3VArp 24-48 \W-f roA A ft-K. <br /> After 48 hours incubation, The(1-20#) are not bacteria growth, Positive(11#) bacteria growth well. <br /> *it Note: (—): *jPfflit represent negative, (+): TR7tt represent positive. <br /> # i^ Detection Conclusion: <br /> ij# Ti13cut BI test of sterility: The results indicated PASSED <br /> &M Test by: 501A AiA Review by: MIX *1693-Q M Report De 0! 08-19 <br /> A* Company%a": <br />