Sterilization certified in line with MPI MEDICOPLAST <br /> DIN EN ISO 13486:2012/EN ISO 11135-1:2014 <br /> Confirmation of Sterilization <br /> Method: VA 7-5-1-6 <br /> Sterilization date: 2019-01-29 <br /> Sterilization lot: 17253 <br /> I <br /> Customer. MT.Derm GmbH <br /> Your delivery note: 18-EB0285,18-EB0339, <br /> Your requisition no.: 18-EB0345,19-EB0239,19-EB0301,19-EB0321,19-EB0340,19-EB0341,19-EB0343 <br /> 19-EB0344 <br /> Our delivery note: 243829 from 2019-01-31 <br /> Item: 9900031 <br /> Sterilisation Pal, MP01 1 d NK <br /> Pr.70 <br /> Quantity: 10 Pal. <br /> Nominal values Actual values <br /> i <br /> Incoming time: min. 12 h 97 h <br /> Preconditioning <br /> Dwell: min. 16 h 34 h <br /> Conditions: 35-45°C 50-80%rel.humidity 35.45°C 50-80%rel.humidity <br /> Sterilization <br /> Cycle: MPI <br /> Temperature: 40±3°C 38`C <br /> Prevacuum: 120±10 mbar abs 127 mbar abs <br /> Hunidity: 50-80%rel.humidity 64,4%rel.humidity <br /> Dwell time: 3 h 3 h <br /> EO concentration (calc.): 1.000±50 mgA 1.000 m911 <br /> Air washes: min. 3 3 <br /> Aerat on <br /> Dwell: min. 1 Days 1 Days <br /> Condition: 35.45°C 35-46 OC <br /> Remarks: <br /> The physical parameters were in accordance to the specifications. <br /> No growth was observed for the microbiological inprocess control with biological <br /> indicators(Bacillus atrophaeus, 3M Attest). <br /> i <br /> The sterilization was conducted in line with the customer validation parameter. <br /> 6RpLqva � 9 <br /> Signature QS: Date: �i 1 !' '� is <br />