Laserfiche WebLink
Sterilization certified in line with rMP1 MEDICOPLAST <br /> DIN EN ISO 13485:2012/EN ISO 11135-1:2014 <br /> Confirmation of Sterilization <br /> Method:VA 7-5-1-6 <br /> Sterilization date: 2019-02-18 <br /> Sterilization lot: 17303 <br /> Customer. MT.Derm GmbH <br /> Your delivery note: 19-EB0639,19-EB0640, <br /> Your requisition no.: 19-EB0649,19-EB0650,19-EB0657,19-EB0659,19-EB0688, <br /> 19-EB0689,19-EB0690,19-EB0691,19-EB0693,19-EB0695 <br /> Our delivery note: 244150 from 2019-02-21 <br /> Item: 9900031 <br /> Sterilisation Pal. MP01 1d NK <br /> Pr.70 <br /> Quantity: 10 Pal. <br /> Nominal values Actual values <br /> Incoming time: min. 12 h 34 h <br /> Precondi#ionina <br /> Dwell: min. 16 h 71,75 h <br /> Conditions: 35-45°C 50-80%rel.humidity 36.45°C 50-80%rel.humidity <br /> Sterilization <br /> Cycle: MP1 <br /> Temperature: 40±3°C 39°C <br /> Prevacuum: 120±10 mbar abs 123 mbar abs <br /> Humidity: 50-80%rei.humidity 64,0°/6 rei.humidity <br /> Dwell time: 3 h 3 h <br /> EO concentration (calc.): 1.000±50 mg/1 1.000 mg/1 <br /> Air washes: min. 3 3 <br /> Aeratlon <br /> Dwell: min 1 Days 2 Days <br /> Condition 35-45°C 35-45 6C <br /> Remarks: <br /> The physical parameters were in accordance to the specifications. <br /> No growth was observed for the microbiological inprocess control with biological <br /> indicators(Bacillus atrophaeus, 3M Attest). O r� <br /> The sterilization was conducted in line with the customer validation parameter. <br /> Approval <br /> �t fFT Signature QS: Date: f ; / 0 <br />