Laserfiche WebLink
�► i <br /> Guixi City Jinlong Tattoo Equipment Manufactory <br /> 3Z#04-1 ' File Number: JLTS-1708 XMtg -k� Chamber No.: U17-08-12/CE <br /> XMttMTVA-V,-M M9ip,3p, ffl H M Test Date: 2017-8-12 <br /> Original Record of BL Growth Test after Sterilization <br /> p$S Customer: Goldentimes Corp <br /> n , Product Name:EZ Cartridges <br /> 11 PIPAO; Product Lot No.:17-08 <br /> XM A X31 Sterilization Date: 17-8-12 TcMaSterilization Lot.: JLWX20170812 <br /> &LMI]tj Test Quantity: 20 pcs *A9g R Test Item: BI Growth Test Record <br /> fag9 H M Test Date: 2017-8-12 *9964 Test Standard: ISO 11737-2:2009 <br /> _ jj ;T; jjf>f(]idjR* BI Growth Test Record Time Table <br /> Strain: Bacillus Atrophaeus ATCC9372 BI Manufacturer: Hangzhou Future Biotech Co., Ltd <br /> BI A-5 Lot No.: 1702 f,"J ifij Incubation Tempearture: 36 V <br /> -Mr` ii� tigjl5rt>f I'0j1Sf-K-,'ffifi Culture Time and BI Growth <br /> Strip No.: 8 16 24 32 40 48 <br /> `-1f�JJ#A T�11 <br /> is 1-10# -- -- -- -- -- -- <br /> 11# + + + + + + <br /> itA Result: (1-10#) V*AKOcMI-rc, 11t1!, (11#) 4EJfA,4 2448/1,1I4)Ai<,o <br /> After 48 hours incubation, The (1-20#) are not bacteria growth, Positive(11#) bacteria growth well. <br /> Note: (--): 3K%ritlfi-t represent negative, (+): T it represent positive. <br /> tA, *9MiQ Detection Conclusion: <br /> IJ#aT$JJ�Ei �: BI test of sterility: 7th A 4 The results indicated PASSED <br /> fi LAI] Test by: V1 Y-1A Review by: k� #bt A�� Report "6:2017-8A 4 <br /> it* Compan tamp: l <br />