Laserfiche WebLink
• F 0 0 <br />CuraCure <br />1350 Lone Palm Avenue, P.O. Box 5237, Modesto, California 95352 (209) 524-6163 <br />May 23, 1988 <br />Mr, Jim Blaise <br />Chief Engineer <br />L. ,.= r%al-Haspital <br />975 South Fairmont Ave. <br />Lodi, CA 95240 <br />AUTOCLAVE CERTIFICATION INSPECTION <br />All steam autoclaves were tested, for proper temperature cali- <br />bration on May 20, 1988. The results of,the tests are as follows: <br />Testing was performed using a Doric Pyrometer, Model 450ET, <br />SN 03545, <br />DMT/lb <br />f;t.t:,r1�f I;,�I V'. `l? 1i)r)c): tiCd(1C� �lOfiiE�7 � ,1 •`:'i '':3 <br />Douglas M. Tayj`br <br />Regional Manager <br />LG"ATION <br />MANUFACTURER <br />SERIAL NUMBER <br />SET POINT <br />ACTUAL <br />Surgery <br />Amsco <br />0129580-016 <br />270°F <br />272.1°F <br />Surgery <br />Amsco <br />0130280-001 <br />270°F <br />270.8°F <br />Surgery <br />C.S. <br />C.S. <br />Amsco <br />Amsco <br />Amsco <br />730504-14 <br />860618-1 <br />680919-2 <br />270°F <br />270°F <br />270°F <br />272.1°F <br />271.60F <br />271.8°F <br />C.S. <br />Amsco <br />0100781-010 <br />270°F <br />272,6°F <br />Laboratory <br />Amsco . <br />309678 <br />2 50 ° <br />F <br />251 -.VF <br />Testing was performed using a Doric Pyrometer, Model 450ET, <br />SN 03545, <br />DMT/lb <br />f;t.t:,r1�f I;,�I V'. `l? 1i)r)c): tiCd(1C� �lOfiiE�7 � ,1 •`:'i '':3 <br />Douglas M. Tayj`br <br />Regional Manager <br />