Laserfiche WebLink
Mr. Loeh.liu Caffey <br /> July 19,2006 <br /> Page 2 <br /> Gas Testing Instrumentation and Methods <br /> Testing for methane gas, oxygen gas,carbon dioxide gas and pressure was performed using a <br /> Lan.dtec Gena.-2000 or. Gastech GT-302 C7as Analyzer calibrated to methane gas. <br /> These instruments measure combustible gas concentrations in air directly on either of two scales, the <br /> first as percent by volume of the lower explosive limit(.LEL) of rnetbane gas in air(0 to 5 percent); <br /> the other as percent by volume (0 to 100 percent) in the gas sampled. The detection limit for both <br /> instruments is approximately 0.1.0 percent by volume. <br /> Gas monitoring well testing procedures are as follows: <br /> • Connect sample tubing, open test valve; observe pressure and.record data. <br /> • For probes 20 feet deep or less, turn on sample pump, observe and record methane, <br /> oxygen, and carbon dioxide gas concentrations, when readings have stabilized for 30 <br /> seconds. <br /> ® For probes more than 20 feet deep, tum on sample pump, extract a minimum.of one <br /> probe volume, observed and record methane,oxygen, and carbon dioxide gas <br /> concentrations, when readings have stabilized for 30 secoz).ds. <br /> ® Close test valve, disconnect sample instruments and secure sample location. <br /> SCS monitored structures by testing random selected locations where methane gas would most <br /> likely accumulate(e.g., electric outlets and junction boxes, conduits, drains, cabinets and closets). <br /> Other required monitoring documentation (e.g., sampling personnel, date, time and meteorological <br /> conditions) is shown on the attached testing data. tables. <br /> Gas 2%l!onitaring Well Testing <br /> In.accordance with,the approved testing schedule, all perimeter GP probes are to be tested at a <br /> -minimum., on a quarterly basis. During this reporting period, all accessible perimeter Gas <br /> Monitoring Probes defined as GP-Nos. l through GP-25 were tested for-the presence of LFG_ Fest <br /> results (shown on Table l) indicated that al 1 GP probes tested exhibited methane gas concentrations <br /> below the.LEL (5 percent by volume) or no methane gas was detected with the exception of GP-11 <br /> and GP-12 (up to 60.0 percent by volume). By May 23, 2006, methane gas concentrations detected <br /> at both probes b.addecreased to below the LEL in response to system adjustments and expansion. <br /> These results indicated that the subject site perimeter probes are in compliance with the <br /> requirements of Title 27 CCR Sections 2091.9.5 and 30931 through 20934. <br />