Laserfiche WebLink
1, <br /> LABORATORY CONTROL SAMPLE DATA REPORT <br /> GC/MS Volatiles <br /> '..� Client Lot #- - _: G3D220308 Work Order #._.: FME4DIAC-LCS Matrix. . . . .. . .. WATER <br /> LCS Lot-Sample#: G3D230000-239 FME4DIAD-LCSD <br /> _. Prep Date.. . . . . : 04/22/03 Analysis Date-_: 04/22/03 <br /> Prep Batch #. . .: 3113239 <br /> Dilution Factor: 1 <br /> SPIKE MEASURED PERCENT <br /> PARAMETER AMOUNT AMOUNT UNITS RECOVERY RPD METHOD <br /> " 1,1-Dichloroethene 20.0 20.4 ug/L 102 SW846 826OB <br /> 20_0 19-6 ug/L 98 3_7 SW846 8260E <br /> Benzene 20.0 21.3 ug/L 107 SW846 8260E <br /> 20_0 20.9 " ug/L 105 1.8 SW846 826OB <br /> Trichloroethene 20.0 20.7 ug/L 104 SW846 826OB <br /> 20.0 21.5 ug/L 108 3.8 SW846 826OB <br /> ,.) Toluene 20-0 22.2 ug/L 111 SW846 826OB <br /> 20.0 21-2 ug/L 106 4.6 SW846 8260B <br /> _) Chlorobenzene 20_0 19.9 ug/L 100 SW846 8260E <br /> pl 20.0 19.8 ug/L 99 0.34 SW846 8260B <br /> PERCENT RECOVERY <br /> SURROGATE RECOVERY LIMITS <br /> Dibromofluoromethane 106 (Bl - 128) <br /> 101 (81 - 128) <br /> 1,2-Dichloroethane-d4 105- (75 - 134) <br /> 105 (75 - 134) <br /> '. .� Toluene-d8 108 (80 - 121) <br /> 104 (80 - 121) <br /> :. � 4-Bromofluorobenzene 103 (80 - 120) <br /> 104 (80 - 120) <br /> i <br /> NOTE(S) : <br /> Calwlafiorc are pmformcd before rounding ro avoid round-off errors in calculated resotu. <br /> Bold prim denotes control paramer= . <br /> , a <br /> ' 7 <br />--=G3O220308 STL-Sacramento(916)373-5600 16 of 61 <br />