Laserfiche WebLink
samples are used to monitor for cross-contamination and laboratory-induced contaminants or <br /> interferences. <br /> 7.4 Standard Reference Materials: Standard reference materials (SRM) are standards <br /> containing certified amounts of metals in soil or sediment. These standards are used for accuracy <br /> and performance checks of FPXRF analyses. SRMs can be obtained from the National Institute of <br /> Standards and Technology (NIST), the U.S. Geological Survey (USGS), the Canadian National <br /> Research Council, and the national bureau of standards in foreign nations. Pertinent NIST SRMs <br /> for FPXRF analysis include 2704, Buffalo River Sediment; 2709, San Joaquin Soil; and 2710 and <br /> 2711, Montana Soil. These SRMs contain soil or sediment from actual sites that has been analyzed <br /> using independent inorganic analytical methods by many different laboratories. <br /> 8.0 SAMPLE COLLECTION, PRESERVATION, AND STORAGE <br /> Sample handling and preservation procedures used in FPXRF analyses should follow the guidelines <br /> in Chapter Three, Inorganic Analytes. <br /> 9.0 QUALITY CONTROL <br /> 9.1 Refer to Chapter One for additional guidance on quality assurance protocols. All field <br /> data sheets and quality control data should be maintained for reference or inspection. <br /> 9.2 Energy Calibration Check: To determine whether an FPXRF instrument is operating <br /> within resolution and stability tolerances, an energy calibration check should be run. The energy <br /> calibration check determines whether the characteristic x-ray lines are shifting, which would indicate <br /> drift within the instrument. As discussed in Section 4.10, this check also serves as a gain check in <br /> the event that ambient temperatures are fluctuating greatly (> 10 to 20°F). <br /> The energy calibration check should be run at a frequency consistent with manufacturers <br /> recommendations. Generally, this would be at the beginning of each working day, after the batteries <br /> are changed or the instrument is shut off, at the end of each working day, and at any other time <br /> when the instrument operator believes that drift is occurring during analysis. A pure element such <br /> as iron, manganese, copper, or lead is often used for the energy calibration check. A manufacturer- <br /> recommended count time per source should be used for the check. <br /> 9.2.1 The instrument manufacturer's manual specifies the channel or kiloelectron <br /> volt level at which a pure element peak should appear and the expected intensity of the peak. <br /> The intensity and channel number of the pure element as measured using the radioactive <br /> source should be checked and compared to the manufacturer's recommendation. If the energy <br /> calibration check does not meet the manufacturer's criteria, then the pure element sample <br /> should be repositioned and reanalyzed. If the criteria are still not met, then an energy <br /> calibration should be performed as described in the manufacturer's manual. With some <br /> FPXRF instruments, once a spectrum is acquired from the energy calibration check, the peak <br /> can be optimized and realigned to the manufacturer's specifications using their software. <br /> 9.3 Blank Samples: Two types of blank samples should be analyzed for FPXRF analysis: <br /> instrument blanks and method blanks. An instrument blank is used to verify that no contamination <br /> exists in the spectrometer or on the probe window. <br /> 9.3.1 The instrument blank can be silicon dioxide, a Teflon block, a quartz block, <br /> "clean" sand, or lithium carbonate. This instrument blank should be analyzed on each working <br /> day before and after analyses are conducted and once per every twenty samples. An <br /> CD-ROM 6200 - 10 Revision 0 <br /> January 1998 <br />