Laserfiche WebLink
10.0 CALIBRATION AND STANDARDIZATION <br /> 10.1 Instrument Calibration: Instrument calibration procedures vary among FPXRF <br /> instruments. Users of this method should follow the calibration procedures outlined in the operator's <br /> manual for each specific FPXRF instrument. Generally, however, three types of calibration <br /> procedures exist for FPXRF instruments: FP calibration, empirical calibration, and the Compton peak <br /> ratio or normalization method. These three types of calibration are discussed below. <br /> 10.2 Fundamental Parameters Calibration: FP calibration procedures are extremely variable. <br /> An FP calibration provides the analyst with a "standard less" calibration. The advantages of FP <br /> calibrations over empirical calibrations include the following: <br /> • No previously collected site-specific samples are required, although <br /> site-specific samples with confirmed and validated analytical results for all <br /> elements present could be used. <br /> • Cost is reduced because fewer confirmatory laboratory results or calibration <br /> standards are required. <br /> However, the analyst should be aware of the limitations imposed on FP calibration by particle <br /> size and matrix effects. These limitations can be minimized by adhering to the preparation <br /> procedure described in Section 7.2. The two FP calibration processes discussed below are based <br /> on an effective energy FP routine and a back scatter with FP (BFP) routine. Each FPXRF FP <br /> calibration process is based on a different iterative algorithmic method. The calibration procedure <br /> for each routine is explained in detail in the manufacturer's user manual for each FPXRF instrument; <br /> in addition, training courses are offered for each instrument. <br /> 10.2.1 Effective Energy FP Calibration: The effective energy FP calibration is <br /> performed by the manufacturer before an instrument is sent to the analyst. Although SSCS <br /> can be used, the calibration relies on pure element standards or SRMs such as those obtained <br /> from NIST for the FP calibration. The effective energy routine relies on the spectrometer <br /> response to pure elements and FP iterative algorithms to compensate for various matrix <br /> effects. <br /> Alpha coefficients are calculated using a variation of the Sherman equation, which <br /> calculates theoretical intensities from the measurement of pure element samples. These <br /> coefficients indicate the quantitative effect of each matrix element on an analyte's measured <br /> x-ray intensity. Next, the Lachance Traill algorithm is solved as a set of simultaneous <br /> equations based on the theoretical intensities. The alpha coefficients are then downloaded <br /> into the specific instrument. <br /> The working effective energy FP calibration curve must be verified before sample <br /> analysis begins on each working day, after every 20 samples are analyzed, and at the end of <br /> sampling. This verification is performed by analyzing either an NIST SRM or an SSCS that is <br /> representative of the site-specific samples. This SRM or SSCS serves as a calibration check. <br /> A manufacturer-recommended count time per source should be used for the calibration check. <br /> The analyst must then adjust the y-intercept and slope of the calibration curve to best fit the <br /> known concentrations of target analytes in the SRM or SSCS. <br /> A percent difference (%D) is then calculated for each target analyte. The %D should <br /> be within ±20 percent of the certified value for each analyte. If the %D falls outside this <br /> acceptance range, then the calibration curve should be adjusted by varying the slope of the <br /> CD-ROM 6200 - 13 Revision 0 <br /> January 1998 <br />