Laserfiche WebLink
by adding metal oxides to a "clean" sand or silicon dioxide matrix that simulates soil. Metal oxides <br /> can be purchased from various chemical vendors. If standards are made on site, a balance capable <br /> of weighing items to at least two decimal places is required. Concentrated ICP or AA standard <br /> solutions can also be used to make standards. These solutions are available in concentrations of <br /> 10,000 parts per million, thus only small volumes have to be added to the soil. <br /> An empirical calibration using SSCS involves analysis of SSCS by the FPXRF instrument and <br /> by a conventional analytical method such as ICP or AA. A total acid digestion procedure should be <br /> used by the laboratory for sample preparation. Generally, a minimum of 10 and a maximum of 30 <br /> well characterized SSCS, site-typical standards, or prepared metal oxide standards are required to <br /> perform an adequate empirical calibration. The number of required standards depends on the <br /> number of analytes of interest and interfering elements. Theoretically, an empirical calibration with <br /> SSCS should provide the most accurate data for a site because the calibration compensates for site- <br /> specific matrix effects. <br /> The first step in an empirical calibration is to analyze the pure element standards for the <br /> elements of interest. This enables the instrument to set channel limits for each element for spectral <br /> deconvolution. Next the SSCS, site-typical standards, or prepared metal oxide standards are <br /> analyzed using a count time of 200 seconds per source or a count time recommended by the <br /> manufacturer. This will produce a spectrum and net intensity of each analyte in each standard. The <br /> analyte concentrations for each standard are then entered into the instrument software; these <br /> concentrations are those obtained from the laboratory, the certified results, or the gravimetrically <br /> determined concentrations of the prepared standards. This gives the instrument analyte values to <br /> regress against corresponding intensities during the modeling stage. The regression equation <br /> correlates the concentrations of an analyte with its net intensity. <br /> The calibration equation is developed using a least squares fit regression analysis. After the <br /> regression terms to be used in the equation are defined, a mathematical equation can be developed <br /> to calculate the analyte concentration in an unknown sample. In some FPXRF instruments, the <br /> software of the instrument calculates the regression equation. The software uses calculated <br /> intercept and slope values to form a multiterm equation. In conjunction with the software in the <br /> instrument, the operator can adjust the multiterm equation to minimize interelement interferences <br /> and optimize the intensity calibration curve. <br /> It is possible to define up to six linear or nonlinear terms in the regression equation. Terms can <br /> be added and deleted to optimize the equation. The goal is to produce an equation with the smallest <br /> regression error and the highest correlation coefficient. These values are automatically computed <br /> by the software as the regression terms are added, deleted, or modified. It is also possible to delete <br /> data points from the regression line if these points are significant outliers or if they are heavily <br /> weighing the data. Once the regression equation has been selected for an analyte, the equation can <br /> be entered into the software for quantitation of analytes in subsequent samples. For an empirical <br /> calibration to be acceptable, the regression equation for a specific analyte should have a correlation <br /> coefficient of 0.98 or greater or meet the DOOs of the project. <br /> In an empirical calibration, one must apply the DQOs of the project and ascertain critical or <br /> action levels for the analytes of interest. It is within these concentration ranges or around these <br /> action levels that the FPXRF instrument should be calibrated most accurately. It may not be possible <br /> to develop a good regression equation over several orders of analyte concentration. <br /> CD-ROM 6200 - 15 Revision 0 <br /> January 1998 <br />