3H ETO STERILIZATION SUMMARY REPORT <br /> START TIME:05/01/2016 <br /> EO STERILIZATION REPORT PATCH NO: L10601205 <br /> 'EVERLASTING( ITEM NAME: Everlasting Brows Microblading Blades OVER TIME: 15/01/2016 <br /> BROWS RE-VALIDATION: 15/01/2019 <br /> In <br /> Temperature in the chamber from a minimum of two locations throughout the <br /> Not used <br /> sterilization cycle;ifused Temperature Min <br /> C. N/A N/A N/A <br /> Precondition Temperature and humidity within the preconditioning area(if used),monitored Temperature C. 50 45 54 Range requested <br /> parameter and recorded from a specified position; Humidity %RH 55 30 65 Range requested <br /> Time of commencement and of removal of load from preconditioning(if used)of <br /> each sterilization load; Preconditioning Time Mins 180 120 300 Range requested <br /> Elapsed time between removal of the sterilization load from preconditioning(if <br /> Transfer Time used)and the commencement of the sterilization cycle; Transfer Time Mins N/A N/A N/A Not used <br /> Indication of the satisfactory operation of the chamber gas circulation system Initial Gas Injection Kpa -38 N/A Information to be <br /> (if used)during gas exposure; -40 completed by supplier <br /> End Gas Injection 50 40 65 <br /> -Exposure Temperature C. 50 45 54 <br /> Temperature and pressure in the chamber throughout the sterilization cycle <br /> -Flushing Pressure:. kPa 50 40 65 parameters required <br /> -Vacuum Pressure -10 -10 -6 in the certificate of <br /> Sterilization sterilization to be <br /> Chamber humidity during conditioning by pressure and/or direct monitoring Exposure Humidity %RH 30-85 29.7 93.5 send to AMD-Ritmed <br /> Range for release of sterile <br /> ((Requested <br /> Parameters for release Evidence that the gaseous ethylene oxide has been admitted to the sterilizer Total Injection Gas product <br /> of Sterile Medical chamber; Weight Kg 50 <br /> Device as per ISO <br /> Standard)) Pressure rise and the quantity of ethylene oxide used or the concentration of <br /> `ethylene oxide in the sterilizer chamber ETO concentrationmg/1 : 520 520 <br /> Conditioning time Exposure Time Mins 480 480 480- <br /> Time,temperature,pressure changes(if any)and/or the operation of the air Air Wash Time 30 15 30 No comment <br /> supply(if used)during aeration. Mins <br /> Aeration Time 3 Information to be <br /> completed by supplier <br /> biological If biological indicators are used in routine monitoring,they shall comply with 8.6. <br /> NOTE 1 See also 8.7. N/A N/A N/A N/A N/A Not used <br /> indicators If chemical indicators are used in routine monitoring,they shall comply with 8.8. <br />