Calibration Procedure: <br />An authorized San -I -Pak service technician will run an Auto Calibration Cycle. <br />During the cycle the chamber is allowed to reach operating temperature and pressure and remain there for <br />20 minutes. This gives the temperature and pressure time to stabilize in the chamber. Twenty minutes after <br />reaching temperature and pressure the machine controller records the temperature. At this point the <br />temperature is compared to a steam table to see how close the temperature is to what the reading should be. <br />This is noted and an automatic correction is made to the temperature reading. All cycles from this point on <br />will use this corrected temperature reading. This reading is also verified using an additional temperature <br />probe connected to a fluke model 51 hand held temp unit that is calibrated and certified annually. <br />Once the calibration is complete the appropriate forms are completed with one copy going to the facility <br />and the others filed in the machine files at San -I -Pak. <br />