WORK ORDER # 90455 <br />Control # <br />1446 <br />Building <br />Lodi Memorial Hospital <br />Description <br />STERILIZER, WASTE <br />Location <br />OUTSIDE <br />Manufacturer <br />Mark Costello <br />Class <br />WST STER <br />Model <br />AS -46 <br />Issue Date <br />5/1/2003 Due 6/30/200 <br />Serial <br />LAAT-975 <br />Assigned to <br />Cost Center <br />8460 Plant Maintenance <br />Risk <br />2 Critical <br />Equipment <br />Notes <br />Instructions <br />Safety <br />Inspection <br />Performance Allow the sterilizer to stablize at the set temperature. Using a calibrated thermometer, verify the <br />Inspection temperature displayed on the chart recorder is within +0/ -5 degrees F. Record temperature accuracy <br />in the service notes. <br />Forward copy of of report to Donna. <br />Preventive CLEAN AND CHECK OVERALL CONDITION <br />Maintenance GREASE DOOR AND LATCH <br />REMOVE DEBRI FROM DRAIN INSIDE VESSEL (DO NOT ENTER!) <br />REMOVE STRAINER AND CLEAN <br />VERIFY STEAM PRESSURE AND RECORD 38 LBS. MINIMUM <br />Parts <br />Replacement <br />Service Date Tech Hr Service Notes <br />6/11/200 Duncan <br />3 <br />1.50 Rec order temp at 270 F. Actual Temp 274 F. Replaced steam <br />inlet nipple, cleaned drain line out. <br />Complete 0 Closed By roehra Date 6/11/2003 Hours <br />