M <br />DATE: April 07, 2011 <br />CUSTOMER: St Joseph's Medical Center <br />ADDRESS: 1909 Cemetery Lane CITY: Stockton <br />STATE: CALIFORNIA 95204 <br />CONTACT: John TEL: (209)943-2000 <br />VESSEL NUMBER: 342 TEMPERATURE ACHIEVED: 278 DEGREE F. <br />PRESSURE ACHIEVED: 38 PSIG. THERMOMETER READING: 278 DEGREE F. <br />VALIDATION PERFORMED BY: Shawn Fallan <br />THIS CALIBRATION WAS PERFORMED BY A FACTORY REPRESENTATIVE TO <br />FACTORY SPECIFICATIONS. <br />COMMENTS: The reference temperature gage is not reading correctly, the thermal air vent <br />needs replaced. The safety cylinder is working but is sticking and should be replaced soon before <br />it sticks all the time. The wheels on the carts need replaced along with the casters. The cart <br />tipper pan on the compactor needs to be repaired or replaced. <br />METHOD OF CALIBRATION: MERCURY THERMOMETER, TAYLOR MODEL 214466-2, <br />RANGE 100 TO 300 DEGREE'S FAHRENHEIT. THERMOMETER WAS PLACED <br />AT THE FRONT OF THE VESSEL AND SUSPENDED AT 5" ABOVE THE FLOOR. <br />THE CHART RECORDER SET POINT WAS CALIBRATED TO THE TEMPERATURE <br />ACHIEVED BY THE MERCURY THERMOMETER. <br />THE STERILIZER SHOULD BE RE -CALIBRATED ON OR BEFORE April 7 2012 <br />1225 E. Wakeham Ave., Santa Ana, Ca. 92705 Phone (888) 596-3985 Fax (714)542-3436 <br />I <br />