C <br />CALIBRATION FORM <br />DATE: February 20, 2014 <br />CUSTOMER: St Joseph's Medical Center <br />ADDRESS: 1909 Cemetery Lane CITY: Stockton <br />STATE: CALIFORNIA 95204 <br />CONTACT: Louie TEL: 209 461-5260 <br />VESSEL NUMBER: 341 TEMPERATURE ACHIEVED: 275 DEGREE F. <br />PRESSURE ACHIEVED: 36 PSIG. THERMOMETER READING: 275 DEGREE F. <br />VALIDATION PERFORMED BY: David Nunez <br />THIS CALIBRATION WAS PERFORMED BY A FACTORY REPRESENTATIVE TO <br />FACTORY SPECIFICATIONS. <br />COMMENTS: All gauges were checked and are within factory settings The autoclave is in <br />good condition with no visible problems I adiusted the tem erature pause on the control panel to <br />read the correct temperature Autoclave has a tot of pitting around the sealing surface of the door <br />METHOD OF CALIBRATION: THERMALMETER, FLUKE MODEL 51/52II, RANGE 100 <br />TO 300 DEGREE'S FAHRENHEIT. THERMALMETER WAS PLACED <br />AT THE FRONT OF THE VESSEL AND SUSPENDED AT 5" ABOVE THE FLOOR. <br />THE CHART RECORDER SET POINT WAS CALIBRATED TO THE TEMPERATURE <br />ACHIEVED BY THE THERMALMETER. <br />THE STERILIZER SHOULD BE RE -CALIBRATED ON OR BEFORE February 20, 2015 <br />