Laserfiche WebLink
lailk <br /> Inspection record form of 131s (biological indicators) used <br /> Recording duration : 2018-09-14-2018-09-16 <br /> Customer name : Product name <br /> JENN BOYD INK Microblading Services and Training Microblading Needles <br /> 0.20mm 14, 0.18mm 14 , 0.18mm 16U Shape <br /> Duration of sterilization : 2018-9-12-2018-9-13 <br /> XMETO-E-GR-001-01 XMETOID-GR-018091233 <br /> Biochemical incubator control number Qty of product(Pcs) Sterilization batch <br /> Name of BI: 3MTM ATTESTM BIOLOGICAL Batch number of BI :2020-06 GR <br /> INDICATORS <br /> BI can be stored until:28-06-2019 Model of 81:1264-S <br /> Time at which the culture began (2018-09-12 04: 01) Culture temperature: 37 f_2 T <br /> Inspection results <br /> ( "+" means positive---culture medium will turn yellow,"_,"means negative---culture medium will remain green) <br /> No. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 <br /> Position Al A3 A7 A9 B3 B8 C1 C9 D2 D7 E1 E3 F5 F9 <br /> 24H - - - - - - - - - - - - - - <br /> 48H - - - - - - - - - - - - -No. 15 16 17 18 19 20 Contr Contr <br /> Position G1 H3 11 13 17 19 ol1 o12 <br /> 24H - - - - - - + + <br /> 48H - - - - - - + + <br /> Conclusion All PCDs were- , and the reference samples were all + and the test results were valid. <br /> Completed by ro Approved by1*3 <br /> Date 2018-9-13 Date 2018-9-13 <br />