Laserfiche WebLink
Sterilization certified in line with MPI M EDICOPLAST <br />EN ISO 13485:20161 EN ISO 11135:2014 <br />Confirmation of Sterilization <br />Method: VA S 1 <br />Sterilization date: 2020-11-03 <br />Sterilization lot: 18566 <br />Customer. MT.DERM GmbH <br />Yore requisition no.: 28102020 <br />G818=20-EB3227,GB10=20-EB3233,GB11=20-EB3240,GB4=20-EB3241 <br />Our delivery note: 252863 from 2020-11-05 <br />tem: 9900031 <br />Sterilisation Pal, MP01 1d NK <br />Pr.70 <br />Quantity: 4 Pal. <br />Nominal values <br />Incoming time: min. 12h <br />Preconditioning <br />Dwell: min. 16 h <br />Conditions: 35-05°C 50.80%rel.humidity <br />Sterilization <br />Cycle: <br />MP1 <br />Temperature: <br />40 t3 °C <br />Prevacuum: <br />120:0 0 mbar abs <br />Humidity: <br />50-80% rel.humidity <br />Dwell time: <br />3 h <br />EO mass: <br />16.32 kg <br />Airwashes: <br />min. 3 <br />e atlon <br />A <br />Dwell: <br />Condition: <br />terri <br />min. 1 Days <br />35-05°C <br />Actual values <br />Oh <br />23,5 h <br />3545°C 50A0%reLhumidity <br />40°C <br />124 mbar abs <br />63,2% rel.humidity <br />3h <br />27,5 kg <br />3 <br />arks: No Incoming time needed, use incoming was before November. <br />The evaluation from the BI's is in process. <br />The physical parameters were in accorciance to the speciflcattons. <br />No growth was observed for am= o gca, inprocess control with biological <br />indicators (Bacillus atrophaeus, 3M Attest). <br />Tire sterilization was conducted in line with the customer validation parameter. <br />�T <br />ffiI <br />Signature 0 <br />1,5 Days <br />35-05°C <br />