Laserfiche WebLink
Geosynte& <br /> consultants <br /> approximately 5 feet bgs. The exact depth of probe installation may vary based on the geology <br /> encountered; soil vapor probes will not be placed within clay layers that could hinder sample <br /> collection. The annular space of the borehole will be filled with filter pack sand from the bottom <br /> of the borehole to 6 inches above the top of the screen, then with granular bentonite, hydrated <br /> bentonite, and cement grout to ground surface (Figure 9). Soil vapor probes will be installed with <br /> a flush-mount completion as indicated in Figure 9. <br /> For installation of the soil vapor probe in a dual-purpose boring, where a grab groundwater <br /> sample is planned to be collected(i.e., 81713-11, 81713-13, 81713-15, and 81713-16), the boring <br /> will be backfilled with cement grout to 6 inches below the bottom of the planned filter pack <br /> (approximately 6 feet bgs) followed by 6 inches of dry granular bentonite. The cement grout will <br /> be allowed to cure for two hours prior to constructing the soil vapor probe as described above. <br /> 4.2.5 Soil Vapor Sampling <br /> Following soil vapor probe installation, soil vapor samples will be collected, as outlined in Table <br /> 1. Soil vapor sampling techniques will include leak testing, shut-in testing, and purging in <br /> general accordance with DTSC guidance, including the Advisory (DTSC 2015) and the <br /> Supplemental Guidance (DTSC 2023), as described in the following sections. Soil vapor probes <br /> will be sampled a minimum of 48 hours following probe installation. <br /> 4.2.5.1 Sampling Equipment <br /> At each soil vapor probe location, the sample train will be set up with dedicated well fittings and <br /> new NylaflowTM tubing. The NylaflowTM tubing will be field screened with a PID for the <br /> presence of VOCs before use at each location. Field instruments (PID, multi-gas detector, and <br /> helium detector) will be calibrated prior to conducting sampling activities. <br /> 4.2.5.2 Shut In Testing <br /> After the aboveground sampling equipment is assembled at each location, a shut-in test will be <br /> conducted to test for leaks in the sample apparatus. A vacuum of approximately 100 inches of <br /> water will be applied to the sampling train, and the vacuum will be monitored for approximately <br /> one minute. If the vacuum dissipates during the shut-in test, the aboveground fittings will be <br /> tightened, and the test will be repeated until the vacuum is maintained for one minute, indicating <br /> a leak-free assembly. This protocol will be repeated after leak testing and before sampling to <br /> ensure that the integrity of the sampling equipment seals has been preserved. <br /> 4.2.5.3 Leak Testing <br /> After each soil gas probe location passes a shut-in test, a leak test will be conducted during vapor <br /> probe purging. Laboratory-grade helium will be used for the leak test to check for ambient air <br /> leakage through the annular seal between the soil vapor probe and the ground surface. A shroud <br /> will be temporarily installed at the ground surface around each well, and helium will be added to <br /> the shroud, with the helium concentration ranging approximately between 20% and 40%. The <br /> concentrations of helium in the shroud and the purge samples will be measured with a field <br /> helium detector. The purge samples will be considered acceptable if the helium concentration is <br /> Investigation Work Plan—UST 817B 9 December 2024 <br />