Laserfiche WebLink
I PEIIRP1200-17 <br />NOTE: To do this, direct the fluid flow into a <br />graduated cylinder while timing for 60 seconds. <br />9. Continue to adjust the size of the test appa- <br />ratus leak orifice until the desired volume is <br />achieved. To expedite calibration, make coarse <br />adjustments by measuring the volume of fluid <br />that corresponds to the 15 -second time interval <br />indicated in Table 9-1. The final calibration <br />of the test apparatus leak orifice must be con- <br />ducted by measuring the appropriate volume of <br />fluid over the full 60 -second timeframe. <br />10. Without adjusting the test apparatus leak ori- <br />fice after it has been properly calibrated in Step <br />9, hang up the dispenser nozzle, allowing the <br />pump to turn off. <br />LL While directing the fluid flow from the leak <br />test apparatus into the test can, observe that the <br />ELLD turns the pump on and pressurizes the <br />line. <br />12. Confirm that the simulated leak condition <br />causes the electronic line leak detector to alarm <br />and/or shut down the pump. <br />FIGURE 9-3. The test port on the shear valve under the <br />dispenser provides a location to simulate a leak in the <br />underground piping system. <br />® pei.org <br />NOTE The ELLD may cycle the pump on/ <br />off several times before alarming or shutting <br />down the pump. <br />13. Record the number of test cycles observed <br />before alarm/shut down occurs. <br />NOTE: When required, perform a 0.2 gph or <br />0.1 gph simulated leak test on ELLDs. Follow <br />the manufacturer's test parameters. The ELLD <br />safety and test set up procedures for 3.0 gph <br />simulated leak test should be followed. <br />9.2.7 Pass/Fail Criteria. If the ELLD has iden- <br />tified the simulated leak by generating an alarm <br />and, if applicable, shutting down the STP, the <br />ELLD passes the test. <br />If the ELLD has not identified the simulated leak <br />by generating an alarm and, if applicable, shutting <br />down the STP, the ELLD fails the test. <br />If the STP does not properly cycle on/off under <br />normal fuel system operating conditions, the ELLD <br />fails the test. <br />NOTE: If the leak detector initially fails the <br />test, repeat the test procedure before declaring <br />the test result as "fail." <br />9.2.8 Restore the System to Operational Con- <br />dition. Upon completion of the test: <br />• Shut off power to the STP and perform lockout/ <br />ragout procedures on the circuit breakers. Allow <br />line pressure to bleed off to zero and close the <br />shear valve. <br />• Remove the test apparatus from the shear valve <br />body and properly reinstall the plug into the <br />shear valve test port. <br />• Reestablish power to the STP, tum dispenser <br />on and confirm that no leaks are present in the <br />system. <br />• Dispense product into the test can to remove any <br />air from the line and confirm that full product <br />flow is achieved. <br />