Laserfiche WebLink
L <br /> L. <br /> June 13, 2006 Lab ID : STK634467 <br /> Neil Anderson & Associates Customer : 3-9574 <br /> V <br /> Quality Control - Organic <br /> Constituent Method Date/II) Type Units Conc. QC Data DQO Note <br /> DBCP 504.1 06/08/2006:A203 Blank ug/L IND <0.01 <br /> LCS ug/L 0.2949 106% 70-130 <br /> BS ug/L 0.2949 101% 70-130 <br /> BSD ug/L 0.2949 105% 70-130 <br /> BSRPD ug/L 4.1% 530 <br /> 504.1 06/08/2006:A 00-CCV ug/L 5.000 97.2% 70-130 <br /> 01-CCV ug/L 5.000 102% 70-130 <br /> 1,3-Dibromopropanc-Surrogate 504.1 06/08/2006:A203 Blank ug/L 0.5868 123% 70-130 <br /> LCS ug/L 0.5868 117% 70-130 <br /> BS ug/L 0.5868 114% 70-130 <br /> BSD ug/L 0.5868 118% 70-130 <br /> 504.1 06/08/2006:A 00-CCV ug/L 9.975 I15% 70-130 <br /> ` 01-CCV ug/L 9.975 118% 70-130 <br /> Definitions <br /> Blank Method Blank - Prepared to verify that the preparation process is not contributing contamination to the samples. <br /> LCS Laboratory Control Standard/Sample - Prepared to verify that the preparation process is not affecting analyte <br /> ■. recovery. <br /> MS/MSD Matrix Spikes - A random sample is spiked with a known amount of analyte. The recoveries are an indication o <br /> how that sample matrix affects analyte recovery. <br /> BSBSD Blank Spikes - A blank is spiked with a known amount of analyte. It is prepared to verify that the preparation <br /> ` process is not affecting analyte recovery. <br /> CCB Continuing Calibration Blank - Analyzed to verify the instrument baseline is within criteria. <br /> CCV Continuing Calibration Verification - Analyzed to verify the instrument calibration is within criteria. <br /> ND Non-detect - Result was below the DQO listed for the analyte. <br /> <'6 High Sample Background - Spike concentration was less than one forth of the sample concentration. <br /> DQO Data Quality Objective - This is the criteria against which the quality control data is compared. <br /> r <br /> V <br /> V <br /> r <br /> STK634467: Quality Control Page 2 <br />