Laserfiche WebLink
to the CGI inlet or blanketing the CGI detector element (element) with an atmosphere from <br /> a calibration gas cylinder. The CGI readout is then set to that value using whatever device (set- <br /> screw, etc.) the CGI possesses to adjust the instrument to the calibration value. <br /> V <br /> Depending on which calibration gas is used for a particular CGI, the instrument manufacturer <br /> normally provides data on the relative response of other organic solvents relative to the <br /> "factory"calibration. For example, a CGI calibrated for methane could be used rto measure <br /> benzene, if the manufacturer supplies a relative response curve for that compound. Of course, <br /> most solvent measurements are of mixtures and it is often difficult to accurately assess these <br /> mixtures. This is one reason that CGI's are normally set to alarm at conservative levels (usually <br /> 20 percent of the LEL of the organic compound chosen as the calibration standard). <br /> 510.3.2 Oxygen Meter <br /> The meter should be allowed to warm up for a few minutes prior to calibration. A cylinder of <br /> calibration gas (typically containing 17 percent OZ and 25,000 ppm methane). is attached to the <br /> meter inlet. (In the case of passive air flow instruments, a hood or cover is placed over the <br /> V <br /> detectors and this in turn is connected to the calibration gas.) Calibration to 21 percent of OZ <br /> may also be performed using ambient air. <br /> V <br /> The calibration gas is turned on and is allowed to stabilize for 15 to 30 seconds. The meters <br /> are then adjusted (following manufacturer's instructions) to proper calibration. This usually <br /> involves adjusting a set screw at the base of the meter. The instrument should then be <br /> rechecked for proper return to zero. <br /> 510.3.3 HNU Photoionization Detector (Model PI 101) <br /> The instrument is turned to the battery check position first. Assuming proper charge, the main <br /> V <br /> control switch is set to the standby position using the zero knob. The instrument is then <br /> allowed to warn up for about five minutes in the standby mode. The calibration gas (usually <br /> isobutylene from the HNU •factory) is attached per the calibration test set up diagram <br /> (attached). The proper range-setting is selected (usually 0 to 200 ppm) and the calibration gas <br /> turned on. <br /> �. Revised: 11/93 HS-510-2 Woodward-Clyde <br />