Laserfiche WebLink
1 <br /> RESNA industries Inc <br /> Quality Assurance/Quality Control Plan <br /> Revised November 1992 <br /> 7.0 FIELD EQUIPMENT CALIBRATION AND MAINTENANCE <br /> 1 The following measuring equipment may be used during the Remedial Investigation. Equipment <br /> is grouped by field activity. Calibration procedures and frequency are listed by piece. <br /> ' Soil Borines and Well Dimensions <br /> 1 Steel and coated cloth tape. Calibration: none, <br /> 1 Water Level Measurements in Wells <br /> Steel surveyors tape. Calibration: manufacturer supplied temperature correction will be <br /> applied as applicable for field conditions. Electrical well sounders. Calibration: check <br /> against steel surveyor's tape_ <br /> 1 <br /> r anic Vanors <br /> Photoionizarion Detector (PID). Calibration: daily field calibration using a isobutylene <br /> standard as per manufacturers instructions. <br /> 1 <br /> Groundwater pH Measurement <br /> ' Digital pH meter. Calibration: standard pH solutions of 4, 7, and 10 will be utilized for <br /> dally field calibration according to manufacturers instructions. <br /> 1 <br /> Electrical Conductivity <br /> ' Electrical Conductivity meter. Calibration: factory-calibrated annually and periodically <br /> calibrated against laboratory prepared standard calibration solution. <br /> Water Temperature <br /> Mercury or digital thermometers. Calibration- factory calibrated once. <br /> QAQClM FNL 5 <br /> 1 <br /> 1 <br />