Laserfiche WebLink
Job Narrative <br /> 720-24805-1 <br /> Comments <br /> No additional comments. <br /> Receipt <br /> All samples were received in good condition within temperature requirements. <br /> GC/MS VOA <br /> Method 8260B:The laboratory control sample(LCSD)for preparation batch#63445 exceeded control limits for the following analytes: <br /> DCDFM. These analytes were biased high in the LCSD and were not detected in the associated samples;therefore,the data have been <br /> reported. <br /> Method 8260B:The laboratory control sample and laboratory control sample duplicate(LCS/LCSD)for preparation batch 63481 <br /> exceeded control limits for the following analytes:chloromethane and dichlorofluoromethane. These analytes were biased high in the <br /> LCS/LCSD and were not detected in the associated samples;therefore,the data have been reported. <br /> Method 8260B:Due to the failure of one of our storage refrigerators,the results for samples M-5313-34.42,TB-121709 are eatimated <br /> because these samples were not stored continously below 6.0 degrees C. <br /> No other analytical or quality issues were noted. <br /> Page 2 of 40 <br />