Laserfiche WebLink
Analysis Report <br /> Lancaster <br /> • Laboratories <br /> Page 3 of 3 <br /> Quality Control Summary <br /> Client Name ChevronTexaco Group Number- 875698 <br /> Reported 11/28/03 at 09 38 AM <br /> Surrogate Qualaty Control <br /> 4168530 95 87 98 88 <br /> 4168531 97 88 97 87 <br /> 4168532 96 89 99 88 <br /> 4168533 96 89 97 88 <br /> Blank 96 90 98 89 <br /> LCS 96 90 98 91 <br /> MS 95 90 97 89 <br /> MSD 97 90 98 91 <br /> Limits 81-120 82-112 85-112 83-113 <br /> Analysis Name 8260 Master Scan (water) <br /> Batch number P033271AA <br /> Dibromofluoromethane 1,2-Dichloroethane-d4 Toluene-d8 4-Bromofluorobenzene <br /> Blank 97 91 97 89 <br /> LCS 96 89 96 90 <br /> MS 96 91 97 90 <br /> MSD 97 90 97 90 <br /> Limits 81-120 82-112 85-112 83-113 <br /> *-Outside of specification <br /> (1) The result for one or both determinations was less than five times the LOQ <br /> (2) The background result was more than four times the spike added <br /> Lancaster Laboratories Inc <br /> MEMBER 2425New Holland Pike <br /> ® PO Box 12425 <br /> Lancaster PA 17605 2425 <br /> 717 656 2300 Fax 717 656 2681 2216 Rev 3110103 <br />