Laserfiche WebLink
Murray & Associates <br /> STANDARD OPERATING PROCEDURE <br /> CONCERNING GROUNDWATER MONITORING <br /> SOP 7 <br /> The HNU Model 101 or a 580 B Photoionization Analyzer shall be used to measure the <br /> ' concentration of trace gases over a range of Iess than 1 ppm to 2,000 ppm by employing the <br /> principle of photoionization for detection The specific instrument used for investigations related <br /> to hydrocarbon contamination should be calibrated for direct readings in parts per million (ppm) <br /> ' volume/volume of benzene Both of these portable field analyzers consists of two components <br /> (1) probe which contains a fan for moving air into the sensor, an ultraviolet light (provides <br /> ionization energy), an ionization chamber and signal amplifier (2) readout assembly which <br /> ' contains a battery, ion chamber bias, meter readout and control panel Specifics of the detection <br /> principle/theory and functions of various components can be found in the manufactures <br /> instruction manual (for example HNU Systems, Inc ) <br /> ' To assure optimum performance, the photoionization analyzer should be calibrated with a <br /> standard gas mixture of known concentration from a pressurized container A daily procedure <br /> ' for calibration involves bringing the probe and readout in close proximity to the calibration gas, <br /> cracking the valve on the tank and checking the instrument reading This provides a useful spot <br /> check for the instrument <br /> A procedure conducted weekly for more accurate calibration of the instrument from a <br /> pressurized container is to connect one side of a "T" to the pressurized container of calibration <br /> gas, another side of the "T" to a rotameter and the third side of the "T" directly to the 8-inch <br /> extension to the photoionization probe (see Figure 2) Crack the valve of the pressurized <br /> container until a slight flow is indicated on the rotameter The instrument draws in the volume <br /> ' of sample required for detection, and the flow in the rotameter indicates an excess of sample <br /> Now adjust the span pot so that the instrument is reading the exact value of the calibration gas <br /> (If the instrument span setting is changed, the instrument should be turned back to the standby <br /> ' position and the electronic zero should be readjusted, if necessary) <br />