Laserfiche WebLink
Instruction / / <br />4.1 Factory Calibration <br />The module has been factory calibrated to alarm as <br />indicated in Table 4.1 or as marked on the calibra- <br />tion tag shipped with the module. <br />4.2 Frequency of Calibration <br />The manufacturer recommends that the calibration <br />of each gas sensor module be verified monthly dur- <br />ing the first three months of operation and then quar- <br />terly. More frequent checks are necessary during <br />periods of extreme humidity and temperature <br />changes. The monitor should have operated con- <br />tinuously (uninterrupted) for at least 24 hours prior <br />to calibration adjustment. <br />4.3 Calibration Process <br />The output signal of the gas sensor module is cali- <br />brated using a calibration gas mixture containing a <br />known concentration of the gas of interest and a <br />balance of air. The concentration of the span gas <br />must be within the full scale of the sensor module <br />and must be equal to the alarm point desired. <br />Calibration requires application of the span gas to <br />the sensor and adjustment of the sensitivity adjust- <br />ment potentiometer. <br />Warning: During calibration the alarm will turn on <br />and remote alarms connected to the alarm relays <br />will be activated. Disable the remote alarm if neces- <br />sary. <br />4.4 Equipment Required <br />The following tools and equipment will be required <br />for calibration: <br />- Jewelers Screwdriver <br />- Calibration Gas <br />- Calibration gas delivery system <br />For accurate calibration use a gas mixture at the <br />required concentration mixed in an air balance, <br />rather than with an inert gas like nitrogen. This gas <br />and the required delivery equipment such as the <br />Model 1200-26 Calibrator is available from Sierra <br />Monitor Corporation. <br />4.5 Calibration Procedure <br />1. The monitor should be in the safe condition prior <br />to calibration (green LED "ON"). <br />Be sure that the area is non -hazardous <br />before proceeding. <br />Disable alarm devices <br />2. On the 20X, remove the cover prior to expos- <br />ing the sensor with gas. On the 200X series <br />(see Figure 1.1 or 1.2) the sensor is in the lower <br />right hand corner of the electronics board and <br />the cover does not need to be removed. <br />3. Apply calibration gas directly to the sensor. <br />4. If the monitor alarms (red LED "ON") within one <br />minute and stops within one minute of the re- <br />moval of the calibration gas, the monitor is in <br />calibration and requires no adjustment. <br />5. If the monitor fails to alarm within one minute <br />of the application of the gas, use a jeweler's <br />screwdriver to adjust the sensitivity adjustment <br />potentiometer (R4, Fig. 4.1) clockwise until the <br />alarm turns "ON". <br />6. If the monitor fails to stop alarming within one <br />minute of removal of the gas, adjust the poten- <br />tiometer counterclockwise until the alarm stops. <br />7. After adjustment, repeat the application and re- <br />moval of gas to verify calibration. <br />8. When the calibration is complete, reconnect <br />any alarm equipment as necessary. <br />Page: 5 20X/200X Gas Sensor Modules (08/04) <br />Calibration <br />Model <br />Time Gas <br />cc/min <br />201/2001 <br />203/2003 <br />206/2006 <br />30 sec. 1000 ppm CH4 <br />1 min. 50 ppm H2S <br />1 min. 100 ppm CO <br />50 <br />50 <br />50 <br />Table 4.1 <br />4.2 Frequency of Calibration <br />The manufacturer recommends that the calibration <br />of each gas sensor module be verified monthly dur- <br />ing the first three months of operation and then quar- <br />terly. More frequent checks are necessary during <br />periods of extreme humidity and temperature <br />changes. The monitor should have operated con- <br />tinuously (uninterrupted) for at least 24 hours prior <br />to calibration adjustment. <br />4.3 Calibration Process <br />The output signal of the gas sensor module is cali- <br />brated using a calibration gas mixture containing a <br />known concentration of the gas of interest and a <br />balance of air. The concentration of the span gas <br />must be within the full scale of the sensor module <br />and must be equal to the alarm point desired. <br />Calibration requires application of the span gas to <br />the sensor and adjustment of the sensitivity adjust- <br />ment potentiometer. <br />Warning: During calibration the alarm will turn on <br />and remote alarms connected to the alarm relays <br />will be activated. Disable the remote alarm if neces- <br />sary. <br />4.4 Equipment Required <br />The following tools and equipment will be required <br />for calibration: <br />- Jewelers Screwdriver <br />- Calibration Gas <br />- Calibration gas delivery system <br />For accurate calibration use a gas mixture at the <br />required concentration mixed in an air balance, <br />rather than with an inert gas like nitrogen. This gas <br />and the required delivery equipment such as the <br />Model 1200-26 Calibrator is available from Sierra <br />Monitor Corporation. <br />4.5 Calibration Procedure <br />1. The monitor should be in the safe condition prior <br />to calibration (green LED "ON"). <br />Be sure that the area is non -hazardous <br />before proceeding. <br />Disable alarm devices <br />2. On the 20X, remove the cover prior to expos- <br />ing the sensor with gas. On the 200X series <br />(see Figure 1.1 or 1.2) the sensor is in the lower <br />right hand corner of the electronics board and <br />the cover does not need to be removed. <br />3. Apply calibration gas directly to the sensor. <br />4. If the monitor alarms (red LED "ON") within one <br />minute and stops within one minute of the re- <br />moval of the calibration gas, the monitor is in <br />calibration and requires no adjustment. <br />5. If the monitor fails to alarm within one minute <br />of the application of the gas, use a jeweler's <br />screwdriver to adjust the sensitivity adjustment <br />potentiometer (R4, Fig. 4.1) clockwise until the <br />alarm turns "ON". <br />6. If the monitor fails to stop alarming within one <br />minute of removal of the gas, adjust the poten- <br />tiometer counterclockwise until the alarm stops. <br />7. After adjustment, repeat the application and re- <br />moval of gas to verify calibration. <br />8. When the calibration is complete, reconnect <br />any alarm equipment as necessary. <br />Page: 5 20X/200X Gas Sensor Modules (08/04) <br />