Laserfiche WebLink
Accuracy is measured by Percent Recovery as in: <br />% recovery = (measured concentration) x 100 <br />(actual concentration) <br />Precision is measured using duplicate tests by Relative Percent Difference <br />(RPD) as in: <br />RPD = (% recovery test I - % recovery test 2) x 100 <br />(% recovery test I + % recovery test 2)/2 <br />Control limits for accuracy (percent recovery) are based on the average, <br />historical percent recovery +/-3 standard deviation units. Control limits <br />for precision (relative percent difference) range from 0 (identical <br />duplicate DCS results) to the average, historical relative percent <br />difference + 3 standard deviation units. In cases where there is not <br />enough historical data, EPA limits or advisory limits are set, with the <br />approval of the Quality Assurance department. <br />IV Analysis Results <br />Test methods may include minor modifications of published EPA Methods such <br />as reporting limits or parameter lists. Reporting limits are adjusted to <br />reflect dilution of the sample, when appropriate. Solid and waste samples <br />are reported on an "as received" basis; i.e., no correction is made for <br />qW moisture content, unless the method requires or the client requests that <br />such correction be made. <br />Results are on the attached data sheets. <br />E <br />nseco <br />CO-0 C --- <br />