Laserfiche WebLink
�® Although different brands of conductivity meters vazy, calibration procedures are similar for ail <br /> of them. A generalized calibration procedure follows: <br /> 1: Turn on the instrument,check the battery,and allow to Warm up if nec. <br /> 27 Rinse the coriductivity probe and temperature probe with dcioaized water then place in <br /> the standard solution, At Forward I conductivities are usually within 500 to 1500c <br /> urnhosA,therefore the 1000 uMhOsA solution is appropriate to use. <br /> 3. Adjust the meter for the appropriate conductivity scale and temperature then read the <br /> conductivity. <br /> 4- If required use.the "caprate" adjustment on the meter to correct.the ' . ent to read <br /> the correct conductivity. <br /> The calibration process will be.recorded on a conductivity calt`bzation-log_ Copies of calibration• <br /> logs for each individual•instrument- arc-kept in,a: ' der.-attthc-office: The logs-are reviewed <br /> Periodically to assess whether the. ' eut is becoming more difficult to calibrate .and <br /> thetafore needs adjustment by the manufacturer,or if new probes are-requi <br />