n <br />reading varies by more than one unit between calibration checks, the meter will be <br />recalibrated. Multiple physical conditions can cause variations in readings, including bubbles <br />in the sampled water, wet or dirty sample containers, a wet or dirty lens, a wet or dirty optical <br />sensor, or leakage of incidental light into the sample chamber. <br />Multiple Sensor Meter (pH Dissolved Oxygen Conductivity Temperature Turbidity): A <br />multiple sensor meter may be used for multiple parameter measurements during sampling. <br />Calibration will be performed prior to initiation of sampling activities, using manufacturer <br />auto -calibration solution. If any of the readings are outside of the manufacturers <br />specifications, the meter will be recalibrated for the parameter outside of the calibration <br />range. Calibration checks will be performed daily. <br />Equipment not listed herein will be calibrated according to manufacturers' recommendations <br />and/or generally accepted practice. Calibration procedures will be documented for the project <br />file. Instruments for which calibration cannot be easily checked will be either tested against <br />another instrument of a similar type, or will be returned to the manufacturer for appropriate <br />calibration. If tested against another instrument capable of making the same measurements, <br />variation between instruments must not exceed five percent. If readings vary more than five <br />percent, the instrument will be returned to the manufacturer for calibration. <br />Scheduled periodic calibration of testing equipment will not relieve field personnel of the <br />responsibility of employing properly functioning equipment. If equipment malfunction is <br />suspected, the device will be removed from service, tagged so that it is not inadvertently used, <br />and the appropriate personnel notified so that re -calibration can be performed or a substitute <br />piece of equipment can be obtained. <br />Equipment Maintenance <br />Maintenance responsibilities for field equipment are coordinated through an instrument <br />technician who is responsible for ensuring that available equipment and instrumentation are <br />ready for use, and that returned equipment is inspected, serviced, and returned to available <br />inventory in a timely manner. Maintenance during use is the responsibility of the field team <br />using the equipment. Calibration logbooks contain information on instrument maintenance, <br />calibration, and repair. A separate logbook is maintained for each instrument. The paperwork <br />will include a detailed listing of the item that was cleaned/replaced, and the make/model/serial <br />number for the particular piece of equipment. <br />5 <br />DAFor rdSAplan.doc <br />