Laserfiche WebLink
April 8, 2014 Lab ID : STK1432472 <br /> Musco Family Olive Environmental Customer : 3-11544 <br /> Quality Control- Organic <br /> Constituent Method WWII) Type Units Conc. QC Data DQO \ote <br /> Organic <br /> TOC 5310C 04/07/14:203672AMM Blank mg/L ND <0.3 <br /> BS mg/L 15.00 103% 75-114 <br /> BSD mg/L 15.00 103% 75-114 <br /> BSRPD mg/L 15.00 0.03% <23.0 <br /> �310C 04/07/14:204889AMM I CCV ppm 1 15.00 104% 67-122 <br /> CCV 22M 15.00 110% 67-122 <br /> Definition <br /> CCV :Continuing Calibration Verification-Analyzed to verify the instrument calibration is within criteria. <br /> Blank :Method Blank-Prepared to verify that the preparation process is not contributing contamination to the samples. <br /> BS :Blank Spikes-A blank is spiked with a known amount of analyte.It is prepared to verify that the preparation process is not <br /> affecting analyte recovery. <br /> BSD :Blank Spike Duplicate of BS/BSD pair-A blank duplicate is spiked with a known amount of analyte.It is prepared to verify that <br /> the preparation process is not affecting analyte recovery. <br /> BSRPD :BS/BSD Relative Percent Difference(RPD)-The BS relative percent difference is an indication of precision for the preparation <br /> and analysis. <br /> ND :Non-detect-Result was below the DQO listed for the analyte. <br /> DQO :Data Quality Objective-This is the criteria against which the quality control data is compared. <br /> Page 1 I of 11 <br />